Why won't some of my P3200 beds not go as low as others?
I have a number of Versacare P3200 beds but some won't go all the way down while others will?
Is this a good question?
I have a number of Versacare P3200 beds but some won't go all the way down while others will?
Is this a good question?
Be sure nothing is triggering the obstacle sensors. See if the blower assembly is not resting on the frame when in lower position. Could be Hilo motors need replacing
Was this answer helpful?
Sometimes the tan aux outlet cable causes the obstacle sensors to trigger.
I have seen the bowers assembly be in a position such it will rest on the frame. You can sometimes see chips in the paint. This will also cause the scale to not work if attempting to weigh the patient when the bed is lowered down.