Skip to main content

Zoll Impact Uni-Vent 731 Battery Compartment Replacement

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.

What you need

  1. Zoll Impact Uni-Vent 731 Battery Compartment Replacement, Battery Compartment Kit: step 1, image 1 of 1
    • Remove the battery compartment cover by unscrewing the (4) Phillips 6-32 X 5/16 screws. This is indicated in red with the number 1.

    • Page 25.

  2. Zoll Impact Uni-Vent 731 Battery Compartment Replacement: step 2, image 1 of 2 Zoll Impact Uni-Vent 731 Battery Compartment Replacement: step 2, image 2 of 2
    • Remove the battery by unscrewing the (4)Phillips 6-32 X 2 ¼ screws and detaching the plug from its locking latch. This is indicated in the second picture in blue with the number 2.

    • Page 25.

  3. Zoll Impact Uni-Vent 731 Battery Compartment Replacement: step 3, image 1 of 1
    • Unscrew the (4) 6-32 X 5/16 Phillips screws to remove the damaged battery compartment case.

    • Page 25.

  4. Zoll Impact Uni-Vent 731 Battery Compartment Replacement: step 4, image 1 of 1
    • Rotate battery compartment to mate with upper and lower case cutouts and press firmly into place. Secure with (4) Phillips 6-32 X 5/16 screws provided.

    • Page 25.

  5. Zoll Impact Uni-Vent 731 Battery Compartment Replacement: step 5, image 1 of 1
    • Re-assemble the battery by connecting its cable to the connector (pull on cable to insure it is locked in place) then tightening the (4) Phillips 6-32 X 2 ¼ screws. This is indicated in blue with the number 5.

    • Page 25.

  6. Zoll Impact Uni-Vent 731 Battery Compartment Replacement: step 6, image 1 of 1
    • Re-assemble the battery compartment cover by tightening the (4) 6-32 X5/16 screws. This is indicated in red with the number 6.

    • Page 25.

  7. Zoll Impact Uni-Vent 731 Battery Compartment Replacement: step 7, image 1 of 1
    • Perform HiPot Testing then perform Functional Test using the RCS to verify the replacement is working properly.

    • Page 25.

Conclusion

To reassemble your device, follow these instructions in reverse order.

Michaela Fromm

Member since: 05/04/20

150 Reputation

1 Guide authored

Team

Cal Poly, Team S1-G5, Zimmerman Spring 2020 Member of Cal Poly, Team S1-G5, Zimmerman Spring 2020

CPSU-ZIMMERMAN-S20S1G5

5 Members

5 Guides authored

0 Comments

Add Comment

View Statistics:

Past 24 Hours: 2

Past 7 Days: 5

Past 30 Days: 19

All Time: 319