Philips Respironics V60 Top Cover Replacement

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.

What you need

  1. Philips Respironics V60 Top Cover Replacement, shut down the ventilator: step 1, image 1 of 2 Philips Respironics V60 Top Cover Replacement, shut down the ventilator: step 1, image 2 of 2
    • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.

    • Select Ventilator Shutdown, which will shut down the ventilator.

    • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

    • Disconnect the ventilator from the AC power.

    • Unplug the ventilator

  2. Philips Respironics V60 Top Cover Replacement: step 4, image 1 of 1
    • Disconnect the internal battery from the power harness connector.

  3. Philips Respironics V60 Top Cover Replacement, Top Cover: step 5, image 1 of 1
    • Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.

    • Then remove the right side panel.

  4. Philips Respironics V60 Top Cover Replacement: step 6, image 1 of 1
    • Remove the 4 screws using a screwdriver.

    • Remove the top cover.

Conclusion

To reassemble your device, follow these instructions in reverse order.