Skip to main content

Newport e500 Exhalation Module Disassembly

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.
  1. Newport e500 Exhalation Module Disassembly, Newport e500 Exhalation Module Disassembly: step 1, image 1 of 1
    • Pull up and remove the flexible rubber exhalation outlet boot (push fit).

  2. Newport e500 Exhalation Module Disassembly: step 2, image 1 of 1
    • Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.

  3. Newport e500 Exhalation Module Disassembly: step 3, image 1 of 1
    • Pull the exhalation module straight up to remove it from the GDU.

Conclusion

To reassemble your device, follow these instructions in reverse order.