Newport e360 Top Cover CVR2101M Replacement

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.

What you need

  1. Newport e360 Top Cover CVR2101M  Replacement, Top Cover CVR2101M: step 1, image 1 of 1
    • Turn off the Ventilator before starting this repair.

    • Shut off or disconnect he air and oxygen gas supplies.

  2. Newport e360 Top Cover CVR2101M  Replacement: step 2, image 1 of 1
    • For your safety put on an electrostatic discharge (ESD) wrist strap.

    • To adviod damage to the ventilator always used standard ESD precaustions while performing any service.

  3. Newport e360 Top Cover CVR2101M  Replacement: step 3, image 1 of 1
    • Rotate the device so that the back face is facing you.

    • The back face will look like the picutre.

  4. Newport e360 Top Cover CVR2101M  Replacement: step 4, image 1 of 1
    • Locate and remove the 6 screws (Philips size 1) that secure the cover.

    • The picture locates 3 screws on the left side. The other 3 screws are found at the same location on the right side.

  5. Newport e360 Top Cover CVR2101M  Replacement: step 5, image 1 of 1
    • Grab the handles on the side and pull the cover away from the rest of the device body.

Conclusion

To reassemble your device, follow these instructions in reverse order.

One other person completed this guide.

Kevin Salceda

Member since: 04/09/20

164 Reputation

1 Guide authored

Team

Cal Poly, Team S7-G21, Paton Spring 2020 Member of Cal Poly, Team S7-G21, Paton Spring 2020

CPSU-PATON-S20S7G21

3 Members

3 Guides authored

0 Comments

Add Comment

View Statistics:

Past 24 Hours: 0

Past 7 Days: 0

Past 30 Days: 1

All Time: 149