How to Perform Circuit Service on the Newport HT70

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.
  1. How to Perform Circuit Service on the Newport HT70, Setup: step 1, image 1 of 1
    • Connect the breathing circuit to the main flow outlet, or as if it were being used for a patient.

    • Cover up the breathing circuit (do not use test lung) .

  2. How to Perform Circuit Service on the Newport HT70, Begin Test: step 2, image 1 of 1
    • Touch the Circuit Check button at the top of the screen.

    • Press ‘Accept’.

    • Uncover the end of the patient circuit pictured in Step 1.

    • Press the ‘Accept’ button to continue with the circuit check.

    • If test passes, the message “Circuit Check PASSED press accept to confirm” will appear on the screen.

    • If the Circuit check does not pass, do not use the ventilator for patient use. Use an alternate Method of ventilation.

    • If the Circuit Check has failed, there may be an issue with the pump assembly, refer to the “Pump Assembly Replacement” guide to proceed with this diagnosis.

Conclusion

To reassemble your device, follow these instructions in reverse order.