How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.
  1. How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff: step 2, image 1 of 1
    • Close the Peak inspiratory Pressure valve by turning the knob fully clockwise.

  2. How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff: step 3, image 1 of 1
    • Adjust the Max Pressure Relief knob clockwise or counterclockwise until manometer reads 40 cmH20

  3. How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff: step 4, image 1 of 1
    • Turn the Peak Inspiratory Pressure knob counterclockwise so the manometer reads 20 cmH20 and shut off the gas flow.