Skip to main content

Hamilton-T1 Battery Replacement

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.
  1. Hamilton-T1 Battery Replacement, Battery: step 1, image 1 of 2 Hamilton-T1 Battery Replacement, Battery: step 1, image 2 of 2
    • Use the handle on the left side of the front panel to pull the cover off.

  2. Hamilton-T1 Battery Replacement: step 2, image 1 of 1
    • Locate the metal lock clip above the battery on the left-hand side.

    • Turn the metal lock clip to the left and all the way up.

    • This will allow access to the battery.

  3. Hamilton-T1 Battery Replacement: step 3, image 1 of 1
    • Pull the white tab on the end of the battery to pull it out of the slot.

    • Insert the replacement battery by holding onto the white tab and sliding it into the empty slot.

    • The Hamilton Medical logo needs to be facing up.

  4. Hamilton-T1 Battery Replacement: step 5, image 1 of 1
    • Turn the metal locking clip down and to the right, until it forms a straight line.

    • Close the cover.

Conclusion

To reassemble your device, follow these instructions in reverse order.