Introduction

This guide was created to assist in either the replacement or maintenance of the various filters contained within the ventilator. Regular checks and maintenance will allow for smooth operation of the ventilator.

NOTES:

  • The side fan filter and rear intake filter can be replaced or washed with soap and water before reuse
  • The HEPA filter must be replaced
  1. The fan filter is located on the side of the ventilator and can be removed and replaced by hand.
    • The fan filter is located on the side of the ventilator and can be removed and replaced by hand.

  2. The air intake filter is located on the rear of the device, and can also be removed by hand.
    • The air intake filter is located on the rear of the device, and can also be removed by hand.

  3. To access the HEPA filter, first remove the screw holding the filter cover in place and gently remove the cover.
    • To access the HEPA filter, first remove the screw holding the filter cover in place and gently remove the cover.

  4. The HEPA filter can now be removed by hand.
    • The HEPA filter can now be removed by hand.

    • Install a new HEPA filter and reattach the back panel.

    • Be sure to test the device and ensure airflow through all filters before use.

Conclusion

To reassemble your device, follow these instructions in reverse order.

Trevor Foley

Member since: 04/10/2020

109 Reputation

1 Guide authored

Team

Cal Poly, Team S11-G24, Paton Spring 2020 Member of Cal Poly, Team S11-G24, Paton Spring 2020

CPSU-PATON-S20S11G24

3 Members

3 Guides authored

0 Comments

Add Comment

View Statistics:

Past 24 Hours: 0

Past 7 Days: 0

Past 30 Days: 7

All Time: 132