Hamilton-C3 Fan Filter Replacement

Introduction

Go to step 1 ↓

Use this guide to replace the cooling fan filter on a Hamilton-C3 ventilator.

The fan filter prevents dust and other coarse particulates from getting blown into and damaging the internals of the ventilator. It should be checked for dust and lint every month (or more often, if necessary) and cleaned or replaced if needed.

A neglected fan filter can cause overheating if it becomes clogged to the point where an insufficient amount of air can be drawn through it by the cooling fan. Inconsistent maintenance of the fan filter can also increase the risk of patient cross-contamination.

For more information on replacing the fan filter, please refer to page 201 in the operator’s manual.

    1. Hamilton-C3 Fan Filter Replacement, Fan Filter: step 1, image 1 of 2 Hamilton-C3 Fan Filter Replacement, Fan Filter: step 1, image 2 of 2
      • Remove the back panel/filter cover (labeled with 3 in the diagram).

      Ask FixBot

      Add Comment

    2. Hamilton-C3 Fan Filter Replacement: step 2, image 1 of 2 Hamilton-C3 Fan Filter Replacement: step 2, image 2 of 2
      • Remove the fan filter.

      • The second image shows what a fan filter should look like.

      • During reassembly, install a new fan filter or wash the existing filter in a mild soap solution, rinse it, dry it, and reinstall it.

      Ask FixBot

      Add Comment

Conclusion

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.

Author

with

Calder Wood

Member since: 04/07/20

270 Reputation

2 Guides authored

Team

Cal Poly, Team S7-G1, Paton Spring 2020 Member of Cal Poly, Team S7-G1, Paton Spring 2020

CPSU-PATON-S20S7G1

3 Members

108 Guides authored

0 Guide Comments

Add Comment

View Statistics:

Past 24 Hours: 5

Past 7 Days: 19

Past 30 Days: 55

All Time: 520