Introduction

This guide will show you how to replace the battery for the Hamilton-C3 Ventilator. There are multiple alarms and indicators that the ventilator uses to signal that its battery is either low or dead. Some of these alarms are identified in the Hamilton-C3 troubleshooting page.


The lithium battery provided by Hamilton Medical is 10.8 V, 6.7 Ah, 72 Wh, and 50 W typical. The typical lifetime of the ventilator without external power running on one battery is about 3.5 hours, and for two batteries 7.5 hours.

For more information on replacing the battery, please refer to page 202 in the operator’s manual.

  1. Connect the ventilator to an external power source. There is a two-prong cable that can be connected to a port on the backside of the ventilator.
    • Connect the ventilator to an external power source. There is a two-prong cable that can be connected to a port on the backside of the ventilator.

    • It is vital that external power is present so that ventilation may continue even without a working battery in place.

  2. Open the battery door on the left side (facing the moniter) of the ventilator. Open the battery door on the left side (facing the moniter) of the ventilator.
    • Open the battery door on the left side (facing the moniter) of the ventilator.

  3. Twist the metal locking clip  (indicated by red arrow) counterclockwise out of the locking position.
    • Twist the metal locking clip (indicated by red arrow) counterclockwise out of the locking position.

    • Use any flathead screwdriver to accomplish this task.

    • During reassembly, turn the metal locking clip (in red) clockwise, back into the locked position, to keep the battery secure.

    • Push the latch (component in red) upwards.

  4. While pushing up the retaining clip, remove the battery (follow the blue arrow, ignore the white arrow).
    • While pushing up the retaining clip, remove the battery (follow the blue arrow, ignore the white arrow).

    • During reassembly, push the new battery in until the latch (circled in red) locks down.

  5. Turn the metal locking clip (in red) clockwise back into the locked position to keep the battery secure.
    • Turn the metal locking clip (in red) clockwise back into the locked position to keep the battery secure.

Conclusion

To reassemble your device, follow these instructions in reverse order.

Lucas Rasmussen

Member since: 04/23/2018

1,904 Reputation

11 Guides authored

Team

Cal Poly, Team S7-G1, Paton Spring 2020 Member of Cal Poly, Team S7-G1, Paton Spring 2020

CPSU-PATON-S20S7G1

3 Members

16 Guides authored

0 Comments

Add Comment

View Statistics:

Past 24 Hours: 0

Past 7 Days: 0

Past 30 Days: 6

All Time: 87