ConMed System 5000 Operator's Manual.pdf
Operator's Manual
LIMITED WARRANTY
For a period of two years following the date of delivery , CONMED Corporation warrants the CONMED System 5000™ Electrosurgical Generator against any defects in material or workmanship and will repair or replace (at CONMED's option) the same without charge, provided that routine maintenance as specified in this manual has been performed using replacement parts approved by CONMED. This warranty is void if the product is used in a manner or for purposes other than intended.
© 2007 CONMED Corporation
525 French Road Utica, New York 13502 U.S.A.
U.S. Patent Nos. 4,961,739 - 5,152,762 - 5,626,5756,830,569 - 6,835,082 - 6,875,210 - 6,939,347 D477,082 - D477,408.
For Technical Service or Return Authorization Phone: 303-699-7600 / 1-800-552-0138 Extension 5274 Fax 303-699-1628
For Customer Service or to order parts phone: 1-800-448-6506 / 315-797-8375 / Fax 315-735-6235 or contact your CONMED Representative.
European Authorized Representative MDSS GmbH Schiffgraben 41 D - 30175 Hannover Germany
The revision level of this manual is specified by the highest revision letter found on either the inside front cover or enclosed errata pages (if any).
Manual Number 60-8016-ENG Rev. N 11/07
Unit Serial Number_________________________________
TM
Table of Contents & List of Illustrations
Table Caption: This table is a table of contents or index for a technical manual, listing sections, their titles, and the corresponding page numbers.
Table Caption: This table is a table of contents or index, listing sections, their titles, and corresponding page numbers. It also includes figures with their titles and page numbers.
TM
General Information Section 1.0
This manual provides the set up and operating instructions for the System 5000™ Electrosurgical Unit (ESU). Electrosurgery can be dangerous to patients, staff and other equipment if misused. Please understand and follow the warnings and cautions that are included in this manual. Technical specifications, performance characteristic curves and user maintenance instructions are also included.
The System 5000™ provides a broad range of capabilities in a single, general-purpose electrosurgical generator. This rugged ESU fulfills the operational and safety needs of the modern operating room by providing:
- Four monopolar cutting modes: Pure, Blend 1, Blend 2 and Blend 3.
- Three monopolar coagulation modes: Spray, Standard and Pinpoint.
- Two bipolar modes: Micro and Macro
- Two specialty modes and a general surgery mode:
- General Mode provides full power performance for open surgical procedures.
- Fluids Specialty Mode provides immediate energy delivery for procedures performed in a fluid medium.
- Laparoscopic Specialty Mode provides optimal safety by limiting output voltage and minimizing the potential harmful effects of capacitive coupling.
- Pulse Cut Mode provides precise modulated energy delivery for critical dissection.
- Pulse Coagulation Mode provides a modulated waveform for unsurpassed precision and control.
- Radio Frequency (RF) isolated and independent outputs.
- The proven Automatic Return Monitor (A.R.M.™) contact quality monitoring system.
- Continuous microprocessor safety monitoring.
Features include:
- Dynamic Response Technology delivers optimal clinical effects in all operational modes through the continuous synchronization of current and voltage.
- Bipolar Output Meter provides visual and audible feedback to surgeon during tubal ligations, vasectomies, and other procedures.
- ReadiPlug™ universal accessory receptacle eliminates the need for foot-controlled adapters.
- Nine programmable memory settings provide set-up convenience.
- Automatic programming restores the ESU to the last settings used.
- Remote Power Control (PC) allows power setting changes using standard hand-controlled pencils.
- Independent power setting available for all modes.
- Ability to change power settings from the control panel while the ESU is activated.
- Two handswitched receptacles and a separate footswitched receptacle enable multiple accessory connections.
- Simultaneous activation in non-contact monopolar coagulation modes.
- Channeled accessory receptacles direct plugs into position, making attachments less cumbersome.
- Illuminated receptacles for greater visibility .
- Integrated operating room control system capability .
- Integrated interface for activation of smoke evacuators and similar devices.
1.1 Cautions
This equipment, in conjunction with connected accessories, is intended to produce high-frequency electrical energy for the controlled destruction of tissue.
Safe and effective electrosurgery is dependent not only on equipment design, but also on factors under the control of the operator. It is important that the instructions supplied with this equipment be read, understood and followed in order to ensure safe and effective use of the equipment.
1.1.1 Cautions For Equipment Preparation
- Use only accessories that comply with the relevant regulatory standards for your location and meet the requirements of Section 1.2, Section 1.4, and Figure 2.4. Use of other accessories may result in increased emissions or decreased immunity of the ESU.
- Reusable accessory cables should be periodically function and safety tested in accordance with the original manufacturer's instructions.
- Visually inspect all accessories before each use to verify the integrity of insulation and the absence of obvious defects. In particular, electrode cables and endoscopic accessories should be checked for damage to the insulation.
- The System 5000™ is equipped to connect three monopolar accessories at one time for the convenience of the surgical staff. Unused accessories should be stowed in a safe, electrically insulated place such as a non-conductive holster, isolated from the patient. CONMED recommends accessories not be connected unless needed.
- Never connect more than one