The LIFEPAK 500 automated external defibrillator (AED) is a portable, battery-powered therapeutic medical device that analyzes the electrocardiograph (ECG) signal of a patient, interprets the indications, advises the operator, and delivers defibrillation therapy.
The LIFEPAK 500 AED records an event log, fault code log, system self-test log, ECG signal channel, and the impedance channel for analysis and record keeping. If the LIFEPAK 500 AED detects a shock able cardiac arrhythmia, it enables the operator to deliver a brief, high-energy pulse of electricity to the heart of the patient. The LIFEPAK 500 AED delivers energy through self-adhesive, disposable defibrillation electrodes. Internal circuity continuously measures the impedance between the electrodes and allows defibrillation only when the defibrillation electrodes are attached to the patient. Advanced arrhythmia analysis controls the critical shock/non-shock decisions.
The LIFEPAK 500 AED is intended for use by personnel who are authorized by a physician/medical director and have, at a minimum, the following skills and training:
- CPR training
- AED training equivalent to that recommended by the American Heart Association
- Training in the use of the LIFEPAK 500 AED
The LIFEPAK 500 AED is intended for use in the hospital and out-of-hospital environments. It has been tested to RTCA/DO-160D, "Environmental Conditions and Test Procedures for Airborne Equipment."
Indications for Use
The LIFEPAK 500 AED is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, not breathing normally, and showing no signs of circulation (for example, no pulse, and/or no coughing, no movement) before the device is used to analyze the patient's ECG rhythm. With Infant/Child Reduced Energy Defibrillation Electrodes, the specially configured biphasic LIFEPAK 500 AED may be used on children who are less than eight years old or who weigh less than 25 kg (55 lb).
Identification
The common sub-assemblies of the LIFEPAK 500 AED are a double-sided surface-mount Main printed circuit board (PCB), a two-line, 20-character, dot matrix Liquid Crystal Display (LCD), a batter, a QUICK-COMBO patient connector serial port, and associate labels, writing, and hardware.
There are two major modes of the LIFEPAK 500 AED, differing in the defibrillation pulse they generate:
- The LIFEPAK 500 Monophasic AED:
- Has a Yellow Exterior with biphasis waveform
- Uses an Energy Storage Capacitor and a wave shaping inductor to deliver a monophasic (Edmark) defibrillation pulse.
- The LIFEPAK 500 Biphasic AED
- Has a Dark Gray exterior with biphasic waveform
- Uses different high-voltage wave shaping components to deliver a biphasic truncated exponential (BTE) defibrillation pulse
Technical Specifications
Physical Characteristics
- Height 10.2 cm (4.0 in)
- Width 26.7 cm (10.5 in)
- Depth 29.5 cm (11.6 in) including handle
- Weight 2.41 kg (5.3 lb) without battery or electrodes (biphasic)
AED
- Input ECG via QUIK-COMBO or FAST-PATCH disposable electrodes; standard placement: anterior-lateral; alternate placement: anterior-posterior
- Electrical Protection Input protected against high voltage defibrillator pulses per IEC 60601/EN 60601
- Safety Classification Internally powered equipment IEC 60601-1/EN 60601-1, 5.1
- Waveform Biphasic and DPS: truncated exponential, with voltage and duration compensation for patient impedance2
- Output Energy Accuracy
- ±10% into 50 ohms (biphasic)
- ±15% into 25 to 100 ohms (biphasic)
- Output Energy Sequence Biphasic and DPS: Three levels, user configurable from 200 to 360 joules, delivered (Level 1, Level 2, Level 3, Level 3...)
- Charge Time With a new, non-rechargeable battery pak, or a new, fully charged rechargeable battery pak:
- 200 joules in less than 9 seconds
- 360 joules in less than 15 seconds
EVENT DOCUMENTATION
- Type Internal digital memory
- Memory Capacity
- 20 minutes audio recording (optional)
- ECG and event log of operator/device actions:
- At least 20 minutes if unit is configured with audio recording and audio recording setup option is ON
- At least 80 minutes if configured with audio recording and audio recording setup option is OFF
- At least 60 minutes if not configured with audio recording
- Report Types CODE SUMMARY report, Event Log report, Test Log report
- Capacity
- 300 Event Log events
- 30 Test Log device tests (assuming no fault codes)
- Communications Options:
- Direct connection to personal computer
- Modem connection to personal computer using Hayes AT-Compatible modem
- Data Review LIFENET system compatible. Options:
- LIFENET DT EXPRESS information management program
- CODE-STAT data management system
ENVIRONMENTAL
- Operating Temperature 0°–50° C* (32°–122° F)
- Storage Temperature
- -30°–65° C* (-22°–149° F) without battery and electrodes
- -30°–65° C* (-22°–149° F) with battery and electrodes, maximum total exposure time limited to one week
- Atmospheric Pressure
- 760 to 429 mmHg (0 to 15,000 ft above sea level)
- DPS: MIL-STD-810E, Method 500.3, Procedure II (Operation): -609.6 m to 4,572 m (-2000 ft to 15,000 ft)
- Relative Humidity 10 to 95% (non-condensing)
- Dust/Water Resistance
- IEC 60529/EN 60529 IPX4 (Splash-proof) with electrodes or connector cover installed
- DPS: IEC 60529/EN 60529 IP54 (Dust/Splash-proof) with electrodes or connector cover installed
- Shock
- MIL-STD-810E, Method 516.4, Procedure 1 (40 g, 6–9 ms pulse, 1/2 sine each axis)
- DPS: IEC 60068-2-29 “Bump” (40 g, 600 bumps)
- Vibration
- MIL-STD-810E, Method 514.4, Helicopter- Category 6 (3.75 g rms) and Ground Mobile - Category 8 (3.15 g rms). RTCA/DO-160C, Table 8-2 Fixed Wing - Turbojet Engine Classification C’ (Fuselage). Test level per Figure 8-5 C. 1 hour in each of three axes.
- DPS: MIL-STD-810F, Method 514.5, Helicopter
- Operational: Sine/Random (1.58 g rms) 1 hour per axis
- Storage: Sine/Random (3.10 g rms) 1 hour per axis
- Jet Aircraft Random (3.54 g rms), 30 min. per axis
- Turboprop Random (4.87 g rms), 1 hour per axis
- Aircraft RTCA/DO-160D (Environmental Conditions and Test Procedures for Airborne Equipment), Section 21, Category M, (Radiated Emissions)
- ESD Electrostatic Discharge: Exceeds EN60601-1-2 (8 kV contact, 15 kV air)
- Salt Fog DPS: MIL-STD-810E, Method 509.3
- EMC Refer to Appendix D
RECHARGEABLE SLA BATTERY PAK
- Type Sealed lead-acid, 8 V, 2.5 amp hours
- Capacity Typical: 59 full discharges or 3 hours of “on time” at 20° C (68° F) with a new, fully charged battery
- Minimum: 43 full discharges with a new, fully charged battery
- Battery Charge Time 10 ±1 hours. Battery charging limited to 15°–35° C (59°–95° F).
- Recommended Replacement Interval Two years or 200 battery charge/discharge cycles, whichever comes first using recommended battery maintenance procedures
- Weight 0.9 kg (1.9 lb)
NON - RECHARGEABLE SLA BATTERY PAK
Lithium sulfur dioxide (LiSO2) battery
- Type Sealed lithium, 12V, 7.5 amp-hours
- Certification
- FAA: TSO-C97 (Battery 3005380-027)
- CAA: BS2G239 (Battery 3005380-028)
- Capacity
- Typical: 312 full discharges or 14 hours of “on time” with a new battery
- Minimum: 230 full discharges with a new battery at 20° C. 0°–58° C (32°–136° F) is a minimum of 197 full discharges with a new battery.
- Shelf Life
- Five years
- Four years (TSO-C97 for aircraft use)
- Weight 0.5 kg (1.2 lb)
Lithium manganese dioxide (LiMnO2) battery
- Type Sealed lithium, 12V, 10 amp-hours
- Certification
- FAA: TSO-C142 (Battery 3201856)
- Capacity
- Typical: 416 full discharges or 18 hours of “on time” with a new battery
- Minimum: 306 full discharges with a new battery
- Shelf Life Five years
- Weight 0.5 kg (1.2 lb)
Additional Information
- Official LIFEPAK 500 Inservice Video (1996)
- Class 2 Device Recall LifePak 500 Automatic external Defibrillator