Background

The Philips HeartStart FRx (model 861304) is a portable automated external defibrillator (AED) used to treat sudden cardiac arrest. After the responder turns it on and places the adhesive pads on the patient's bare chest, the FRx analyzes the heart rhythm and decides whether a shock is needed. If it is, the device charges and tells the responder to press the flashing orange Shock button. Spoken prompts, a CPR metronome and lighted icons guide the responder through each step. The FRx uses one set of pads, SMART Pads II, for both adults and children. Inserting the optional Infant/Child Key lowers the shock energy and switches the voice prompts to infant and child instructions.

Philips Medical Systems received FDA 510(k) clearance for the FRx (K050004) on February 25, 2005. Its FDA submission lists the earlier HeartStart OnSite AED as its comparison (predicate) device. According to its owner's manual, the FRx is intended for responders trained in basic life support, such as firefighters, police officers, flight attendants, teachers, coaches and members of workplace or campus response teams. A Laerdal-branded version, the Laerdal FRx (model 861305), was also sold. Philips gives the FRx an expected service life of 10 years.

Owners can handle most upkeep themselves: checking the green Ready light, replacing the battery and pads before they expire, and responding to chirps or a flashing i-button. The FRx runs automatic self-tests and alerts the owner when something needs attention. The owner's manual states that the FRx has no user-serviceable components and should not be opened. A unit that repeatedly fails its self-test, gives triple chirps, or is physically damaged should be taken out of service and returned to Philips or an authorized service center.

Identification

You can identify the HeartStart FRx by the model number printed on the device label:

  • 861304: Philips HeartStart FRx
  • 861305: Laerdal FRx

The Philips HeartStart FRx is a small, handheld automated external defibrillator in a water and dust resistant case. It measures about 2.4 inches high, 7.1 inches deep and 8.8 inches wide, and weighs about 3.5 pounds with the battery and pads installed. The case is gray-blue with a deeper blue front panel. A brighter blue center panel shows an illustration of where to place the pads on a patient’s chest, with “HEARTSTART FRx” printed above it and “PHILIPS” below it.

The front panel has a green On/Off button, a blue i-button for information and alerts, an orange Shock button marked with a lightning bolt, and a green Ready light that blinks when the unit has passed its last self-test. The Infant/Child Key slot is at the top center of the front panel, and the key itself is partly pink. The M5070A battery installs in a compartment on the back. The SMART Pads II stay sealed in a case, and their cable plugs into the connector port on the FRx.

The HeartStart OnSite (HS1) and HeartStart Home look similar but load their pads in a cartridge with a pull handle on the front, and they have no Infant/Child Key slot. They use separate infant/child pads cartridges instead.

The HeartStart FR2+ and FR3 are aimed mainly at professional responders and use different batteries. Some FR3 models also have a screen that shows the heart rhythm.

Technical specifications

Device Type: Automated External Defibrillator (AED)
Model Numbers: 861304 (Philips), 861305 (Laerdal)
Height: 6 cm (2.4 in)
Width: 22 cm (8.8 in)
Depth: 18 cm (7.1 in)
Weight: 1.6 kg (3.5 lb) (With the battery and pads installed)
Defibrillator Pads: SMART Pads II (989803139261), disposable, self-adhesive pads for adults and children. Each pad is 80 cm², and the pad cable is 1.22 meters (48 in) long
Defibrillator Waveform: SMART Biphasic (impedance-compensated biphasic truncated exponential) waveform
Energy Output: 150 J for adults; 50 J with the Infant/Child Key (989803139311) for children under 25 kg (55 lb) or under 8 years old
Battery: M5070A, 9 VDC lithium manganese dioxide, non-rechargeable
Battery Life: Shelf life of at least 5 years before installation; standby life of typically 4 years once installed
Environmental Temperature: 32° F to 122° F (0° C to 50° C)
Water and Dust Resistance: IP55
Automated Self-Tests: Daily, weekly and monthly, plus a self-test each time a battery is installed
Data Transfer: Infrared port for HeartStart event review software
Expected Service Life: 10 years