{"wikiid":430643,"langid":"en","namespace":"CATEGORY","title":"O-Two CAREvent","revisionid":2553273,"contents_raw":"== Background and Identification ==\n\nThe O-Two CAREvent is a medical ventilator typically used in ATV+ and MRI settings. The ventilator is produced by O-Two Medical Technologies and is intended to assist patients who are having trouble breathing, primarily due to trauma.\n\n* O-Two Medical Technologies have produced state-of-the-art products since 1971.\n* O-Two Medical exports to over 60 different countries and possesses more than 40 patents.\n* The CAREvent ATV+ and MRI consist of a portable control module, input hose and optional single use (or reusable where available) patient circuit. The control groupings are designed to aid in the selection and setting of the breathing parameters. The groupings are colored on the controls to aid in ease of use.\n* The CAREvent ATV+ and MRi are for use in pre hospitals, intra hospitals, interhopsitals, and air ambulance settings. They are designed to be used by suitably trained personnel. The device offers various modes that support the transportation and resuscitation of a variety of patients who require ventilatory support.\n* The ATV+ and MRI are time-cycled which provides a range of frequencies for ventilation. They also include Minute Volume settings to provide a range of delivered tidal volumes and ventilation rates.\n\n== Technical Specifications ==\n\n* ***Dimensions***\n** 9.27 x 7.6 x 4.4 Inches\n** 236 x 194 x 122 mm\n* ***Weight***\n** 2.84kg approximately for the ATV+\n** 3.0kg approximately for the MRI\n* ***Minute Volume Range***\n** 2 - 14 L\/min\n* ***Frequency Range***\n** 8 - 40 BPM\n* ***I:E Ratio***\n** 1:2\n* ***Input Pressure***\n** 45 - 87 PSI (3 - 6 Bar)\n* ***Automatic Flow Rate***\n** 6 - 42 L\/min\n* ***Manual Flow Rate***\n** 6 - 42 L\/min\n* ***Manual Override Delay Time***\n** 17 - 23 Seconds\n* ***Demand Breathing Flow Rate***\n** >100 L\/min @ -6 cmH2O {hPa}\n* ***Demand Breathing Triggering Pressure***\n** -2 cmH2O max\n* ***Auto Shut Off Delay Time***\n** 5-8 Seconds\n* ***CPAP\/PEEP***\n** 0 - 20 cmH2O\n* ***Oxygen Concentration***\n** 60 OR 100%\n* ***Inspiratory\/Expiratory Resistance***\n** <6cmH2O (hPa) @ 60 L\/min\n* ***Operating Temperature***\n** -18C to + 50C\n** 0F to +122F\n* ***Storage Temperature***\n** +40C to +60C\n** -40F to +140F\n* ***Gas Supply Status Indicator***\n** Provides a visual indication of Gas Supply Status\n* ***Pressure Relief Range***\n** 20 - 60 cmH2O\n** 19.6 TO 58.8 mBar\n* ***Low Input Pressure Alarm***\n** Audible pulsed alarm to indicate input pressure is dropping below the minimum requirement. Cannot be switched off\n* ***Airway Pressure Alarm***\n** Audible alarm to indicate that the selected peak airway pressure has been reached\n* ***BSI Alarm***\n** Audible high frequency pulsed alarm indicating that the patient circuit is disconnected or insufficient ventilation pressure (<10 Cmh2o) has been attained\n* ***Pressure Gauge Accuracy***\n** \u00b1 2% Full Scale\n* ***Input Connection***\n** 9\/16\u201d DISS\n* ***Patient Connection***\n** 15\/22 mm\n* ***Patient Valve Dead Space***\n** 8 ml\n* ***Cylinder Duration (Aluminum \u201cD\u201d size cylinder containing 415 litres of oxygen)***\n** On 100% Settings\n*** 2 Litre Mv\n**** 207 Minutes\n*** 14 Litre Mv\n**** 30 Minutes\n** On 60% Settings\n*** 2 Litre Mv\n**** 345 Minutes\n*** 14 Litre Mv\n**** 50 Minutes\n\n== Additional Information ==\n\n[https:\/\/otwo.com\/products\/automatic-transport-ventilators\/hand-held-automatic-ventilators-resuscitators\/carevent-ventilators\/|CAREven Official Page]\n\n[document|2572276|CAREvent User Manual]\n\n[document|2572999|CAREvent Specification Sheet]","contents_json":{"type":"doc","content":[{"type":"paragraph","content":[]},{"type":"heading","attrs":{"level":2},"content":[{"type":"text","text":"Background and Identification"}]},{"type":"paragraph","content":[{"type":"text","text":"The O-Two CAREvent is a medical ventilator typically used in ATV+ and MRI settings. The ventilator is produced by O-Two Medical Technologies and is intended to assist patients who are having trouble breathing, primarily due to trauma."}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"O-Two Medical Technologies have produced state-of-the-art products since 1971."}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"O-Two Medical exports to over 60 different countries and possesses more than 40 patents."}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"The CAREvent ATV+ and MRI consist of a portable control module, input hose and optional single use (or reusable where available) patient circuit. The control groupings are designed to aid in the selection and setting of the breathing parameters. The groupings are colored on the controls to aid in ease of use."}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"The CAREvent ATV+ and MRi are for use in pre hospitals, intra hospitals, interhopsitals, and air ambulance settings. They are designed to be used by suitably trained personnel. The device offers various modes that support the transportation and resuscitation of a variety of patients who require ventilatory support."}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"The ATV+ and MRI are time-cycled which provides a range of frequencies for ventilation. They also include Minute Volume settings to provide a range of delivered tidal volumes and ventilation rates."}]}]}]},{"type":"heading","attrs":{"level":2},"content":[{"type":"text","text":"Technical Specifications"}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Dimensions","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"9.27 x 7.6 x 4.4 Inches"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"236 x 194 x 122 mm"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Weight","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"2.84kg approximately for the ATV+"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"3.0kg approximately for the MRI"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Minute Volume Range","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"2 - 14 L\/min"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Frequency Range","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"8 - 40 BPM"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"I:E Ratio","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"1:2"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Input Pressure","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"45 - 87 PSI (3 - 6 Bar)"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Automatic Flow Rate","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"6 - 42 L\/min"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Manual Flow Rate","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"6 - 42 L\/min"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Manual Override Delay Time","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"17 - 23 Seconds"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Demand Breathing Flow Rate","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":">100 L\/min @ -6 cmH2O {hPa}"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Demand Breathing Triggering Pressure","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"-2 cmH2O max"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Auto Shut Off Delay Time","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"5-8 Seconds"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"CPAP\/PEEP","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"0 - 20 cmH2O"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Oxygen Concentration","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"60 OR 100%"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Inspiratory\/Expiratory Resistance","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"<6cmH2O (hPa) @ 60 L\/min"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Operating Temperature","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"-18C to + 50C"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"0F to +122F"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Storage Temperature","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"+40C to +60C"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"-40F to +140F"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Gas Supply Status Indicator","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Provides a visual indication of Gas Supply Status"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Pressure Relief Range","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"20 - 60 cmH2O"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"19.6 TO 58.8 mBar"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Low Input Pressure Alarm","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Audible pulsed alarm to indicate input pressure is dropping below the minimum requirement. Cannot be switched off"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Airway Pressure Alarm","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Audible alarm to indicate that the selected peak airway pressure has been reached"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"BSI Alarm","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Audible high frequency pulsed alarm indicating that the patient circuit is disconnected or insufficient ventilation pressure (<10 Cmh2o) has been attained"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Pressure Gauge Accuracy","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"\u00b1 2% Full Scale"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Input Connection","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"9\/16\u201d DISS"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Patient Connection","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"15\/22 mm"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Patient Valve Dead Space","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"8 ml"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Cylinder Duration (Aluminum \u201cD\u201d size cylinder containing 415 litres of oxygen)","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"On 100% Settings"}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"2 Litre Mv"}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"207 Minutes"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"14 Litre Mv"}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"30 Minutes"}]}]}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"On 60% Settings"}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"2 Litre Mv"}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"345 Minutes"}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"14 Litre Mv"}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"50 Minutes"}]}]}]}]}]}]}]}]}]},{"type":"heading","attrs":{"level":2},"content":[{"type":"text","text":"Additional Information"}]},{"type":"paragraph","content":[{"type":"text","marks":[{"type":"link","attrs":{"href":"https:\/\/otwo.com\/products\/automatic-transport-ventilators\/hand-held-automatic-ventilators-resuscitators\/carevent-ventilators\/","target":false}}],"text":"CAREven Official Page"}]},{"type":"paragraph","content":[{"type":"text","text":"CAREvent User Manual","marks":[{"type":"siteLink","attrs":{"kind":"document","id":"2572276","options":"","defaultText":null,"newWindow":false,"href":"\/Document\/CSUeFcPVO1ESPnVQ\/O-Two-CAREvent-Automatic-Transport-Ventilator-User-Manual.pdf"}}]}]},{"type":"paragraph","content":[{"type":"text","text":"CAREvent Specification Sheet","marks":[{"type":"siteLink","attrs":{"kind":"document","id":"2572999","options":"","defaultText":null,"newWindow":false,"href":"\/Document\/EkwePYRmOcHExBDx\/O-Two-CAREvent-ATV-Plus-MRI-Specification-Sheet.pdf"}}]}]}]},"contents_rendered":"<div class=\"clearer\"><\/div><div class=\"headerContainer\"><a class=\"selfLink\" href=\"#Section_Background_and_Identification\" title=\"Link to this section\" aria-hidden=\"true\" tabindex=\"-1\"><span    class=\"fa-svg-icon\"><svg width=\"16\" height=\"16\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 576 512\" overflow=\"visible\"><!--! Font Awesome Pro 7.3.1 by @fontawesome - https:\/\/fontawesome.com License - https:\/\/fontawesome.com\/license (Commercial License) Copyright 2026 Fonticons, Inc. --><path d=\"M427.5 88c-21.4 0-42.2 6.9-59.3 19.4-11.4-11.7-24.4-22.1-38.4-30.7 27-23.6 61.7-36.7 97.7-36.7 82 0 148.5 66.5 148.5 148.5 0 39.4-15.6 77.1-43.5 105l-79.1 79.1c-27.8 27.8-65.6 43.5-105 43.5-82 0-148.5-66.5-148.5-148.5 0-17.9 3.2-35.5 9.4-51.9 4.6-12.4 18.5-18.7 30.9-14.1S259 220 254.4 232.4c-4.2 11.1-6.4 23-6.4 35.2 0 55.5 45 100.5 100.5 100.5 26.6 0 52.2-10.6 71-29.4l79.1-79.1C517.4 240.6 528 215.1 528 188.5 528 133 483 88 427.5 88zm-200 56c-26.6 0-52.2 10.6-71 29.4L77.4 252.5c-18.8 18.8-29.4 44.4-29.4 71 0 55.5 45 100.5 100.5 100.5 21.4 0 42.2-6.9 59.3-19.4 11.4 11.8 24.4 22.1 38.4 30.7-27 23.6-61.7 36.7-97.7 36.7-82 0-148.5-66.5-148.5-148.5 0-39.4 15.6-77.1 43.5-105l79.1-79.1c27.8-27.8 65.6-43.5 105-43.5 82 0 148.5 66.5 148.5 148.5 0 17.9-3.2 35.5-9.4 51.9-4.6 12.4-18.5 18.7-30.9 14.1S317 292 321.6 279.6c4.2-11.1 6.4-23 6.4-35.1 0-55.5-45-100.5-100.5-100.5z\"\/><\/svg><\/span><\/a><h2 id=\"Section_Background_and_Identification\" class=\"header\">Background and Identification<\/h2><\/div>\n\n<p>The O-Two CAREvent is a medical ventilator typically used in ATV+ and MRI settings. The ventilator is produced by O-Two Medical Technologies and is intended to assist patients who are having trouble breathing, primarily due to trauma.<\/p>\n\n<ul><li>O-Two Medical Technologies have produced state-of-the-art products since 1971.<\/li><li>O-Two Medical exports to over 60 different countries and possesses more than 40 patents.<\/li><li>The CAREvent ATV+ and MRI consist of a portable control module, input hose and optional single use (or reusable where available) patient circuit. The control groupings are designed to aid in the selection and setting of the breathing parameters. The groupings are colored on the controls to aid in ease of use.<\/li><li>The CAREvent ATV+ and MRi are for use in pre hospitals, intra hospitals, interhopsitals, and air ambulance settings. They are designed to be used by suitably trained personnel. The device offers various modes that support the transportation and resuscitation of a variety of patients who require ventilatory support.<\/li><li>The ATV+ and MRI are time-cycled which provides a range of frequencies for ventilation. They also include Minute Volume settings to provide a range of delivered tidal volumes and ventilation rates.<\/li><\/ul>\n\n<div class=\"clearer\"><\/div><div class=\"headerContainer\"><a class=\"selfLink\" href=\"#Section_Technical_Specifications\" title=\"Link to this section\" aria-hidden=\"true\" tabindex=\"-1\"><span    class=\"fa-svg-icon\"><svg width=\"16\" height=\"16\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 576 512\" overflow=\"visible\"><!--! Font Awesome Pro 7.3.1 by @fontawesome - https:\/\/fontawesome.com License - https:\/\/fontawesome.com\/license (Commercial License) Copyright 2026 Fonticons, Inc. --><path d=\"M427.5 88c-21.4 0-42.2 6.9-59.3 19.4-11.4-11.7-24.4-22.1-38.4-30.7 27-23.6 61.7-36.7 97.7-36.7 82 0 148.5 66.5 148.5 148.5 0 39.4-15.6 77.1-43.5 105l-79.1 79.1c-27.8 27.8-65.6 43.5-105 43.5-82 0-148.5-66.5-148.5-148.5 0-17.9 3.2-35.5 9.4-51.9 4.6-12.4 18.5-18.7 30.9-14.1S259 220 254.4 232.4c-4.2 11.1-6.4 23-6.4 35.2 0 55.5 45 100.5 100.5 100.5 26.6 0 52.2-10.6 71-29.4l79.1-79.1C517.4 240.6 528 215.1 528 188.5 528 133 483 88 427.5 88zm-200 56c-26.6 0-52.2 10.6-71 29.4L77.4 252.5c-18.8 18.8-29.4 44.4-29.4 71 0 55.5 45 100.5 100.5 100.5 21.4 0 42.2-6.9 59.3-19.4 11.4 11.8 24.4 22.1 38.4 30.7-27 23.6-61.7 36.7-97.7 36.7-82 0-148.5-66.5-148.5-148.5 0-39.4 15.6-77.1 43.5-105l79.1-79.1c27.8-27.8 65.6-43.5 105-43.5 82 0 148.5 66.5 148.5 148.5 0 17.9-3.2 35.5-9.4 51.9-4.6 12.4-18.5 18.7-30.9 14.1S317 292 321.6 279.6c4.2-11.1 6.4-23 6.4-35.1 0-55.5-45-100.5-100.5-100.5z\"\/><\/svg><\/span><\/a><h2 id=\"Section_Technical_Specifications\" class=\"header\">Technical Specifications<\/h2><\/div>\n\n<ul><li><strong>Dimensions<\/strong><ul><li>9.27 x 7.6 x 4.4 Inches<\/li><li>236 x 194 x 122 mm<\/li><\/ul><\/li><li><strong>Weight<\/strong><ul><li>2.84kg approximately for the ATV+<\/li><li>3.0kg approximately for the MRI<\/li><\/ul><\/li><li><strong>Minute Volume Range<\/strong><ul><li>2 - 14 L\/min<\/li><\/ul><\/li><li><strong>Frequency Range<\/strong><ul><li>8 - 40 BPM<\/li><\/ul><\/li><li><strong>I:E Ratio<\/strong><ul><li>1:2<\/li><\/ul><\/li><li><strong>Input Pressure<\/strong><ul><li>45 - 87 PSI (3 - 6 Bar)<\/li><\/ul><\/li><li><strong>Automatic Flow Rate<\/strong><ul><li>6 - 42 L\/min<\/li><\/ul><\/li><li><strong>Manual Flow Rate<\/strong><ul><li>6 - 42 L\/min<\/li><\/ul><\/li><li><strong>Manual Override Delay Time<\/strong><ul><li>17 - 23 Seconds<\/li><\/ul><\/li><li><strong>Demand Breathing Flow Rate<\/strong><ul><li>&gt;100 L\/min @ -6 cmH2O {hPa}<\/li><\/ul><\/li><li><strong>Demand Breathing Triggering Pressure<\/strong><ul><li>-2 cmH2O max<\/li><\/ul><\/li><li><strong>Auto Shut Off Delay Time<\/strong><ul><li>5-8 Seconds<\/li><\/ul><\/li><li><strong>CPAP\/PEEP<\/strong><ul><li>0 - 20 cmH2O<\/li><\/ul><\/li><li><strong>Oxygen Concentration<\/strong><ul><li>60 OR 100%<\/li><\/ul><\/li><li><strong>Inspiratory\/Expiratory Resistance<\/strong><ul><li>&lt;6cmH2O (hPa) @ 60 L\/min<\/li><\/ul><\/li><li><strong>Operating Temperature<\/strong><ul><li>-18C to + 50C<\/li><li>0F to +122F<\/li><\/ul><\/li><li><strong>Storage Temperature<\/strong><ul><li>+40C to +60C<\/li><li>-40F to +140F<\/li><\/ul><\/li><li><strong>Gas Supply Status Indicator<\/strong><ul><li>Provides a visual indication of Gas Supply Status<\/li><\/ul><\/li><li><strong>Pressure Relief Range<\/strong><ul><li>20 - 60 cmH2O<\/li><li>19.6 TO 58.8 mBar<\/li><\/ul><\/li><li><strong>Low Input Pressure Alarm<\/strong><ul><li>Audible pulsed alarm to indicate input pressure is dropping below the minimum requirement. Cannot be switched off<\/li><\/ul><\/li><li><strong>Airway Pressure Alarm<\/strong><ul><li>Audible alarm to indicate that the selected peak airway pressure has been reached<\/li><\/ul><\/li><li><strong>BSI Alarm<\/strong><ul><li>Audible high frequency pulsed alarm indicating that the patient circuit is disconnected or insufficient ventilation pressure (&lt;10 Cmh2o) has been attained<\/li><\/ul><\/li><li><strong>Pressure Gauge Accuracy<\/strong><ul><li>\u00b1 2% Full Scale<\/li><\/ul><\/li><li><strong>Input Connection<\/strong><ul><li>9\/16\u201d DISS<\/li><\/ul><\/li><li><strong>Patient Connection<\/strong><ul><li>15\/22 mm<\/li><\/ul><\/li><li><strong>Patient Valve Dead Space<\/strong><ul><li>8 ml<\/li><\/ul><\/li><li><strong>Cylinder Duration (Aluminum \u201cD\u201d size cylinder containing 415 litres of oxygen)<\/strong><ul><li>On 100% Settings<ul><li>2 Litre Mv<ul><li>207 Minutes<\/li><\/ul><\/li><li>14 Litre Mv<ul><li>30 Minutes<\/li><\/ul><\/li><\/ul><\/li><li>On 60% Settings<ul><li>2 Litre Mv<ul><li>345 Minutes<\/li><\/ul><\/li><li>14 Litre Mv<ul><li>50 Minutes<\/li><\/ul><\/li><\/ul><\/li><\/ul><\/li><\/ul>\n\n<div class=\"clearer\"><\/div><div class=\"headerContainer\"><a class=\"selfLink\" href=\"#Section_Additional_Information\" title=\"Link to this section\" aria-hidden=\"true\" tabindex=\"-1\"><span    class=\"fa-svg-icon\"><svg width=\"16\" height=\"16\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 576 512\" overflow=\"visible\"><!--! Font Awesome Pro 7.3.1 by @fontawesome - https:\/\/fontawesome.com License - https:\/\/fontawesome.com\/license (Commercial License) Copyright 2026 Fonticons, Inc. --><path d=\"M427.5 88c-21.4 0-42.2 6.9-59.3 19.4-11.4-11.7-24.4-22.1-38.4-30.7 27-23.6 61.7-36.7 97.7-36.7 82 0 148.5 66.5 148.5 148.5 0 39.4-15.6 77.1-43.5 105l-79.1 79.1c-27.8 27.8-65.6 43.5-105 43.5-82 0-148.5-66.5-148.5-148.5 0-17.9 3.2-35.5 9.4-51.9 4.6-12.4 18.5-18.7 30.9-14.1S259 220 254.4 232.4c-4.2 11.1-6.4 23-6.4 35.2 0 55.5 45 100.5 100.5 100.5 26.6 0 52.2-10.6 71-29.4l79.1-79.1C517.4 240.6 528 215.1 528 188.5 528 133 483 88 427.5 88zm-200 56c-26.6 0-52.2 10.6-71 29.4L77.4 252.5c-18.8 18.8-29.4 44.4-29.4 71 0 55.5 45 100.5 100.5 100.5 21.4 0 42.2-6.9 59.3-19.4 11.4 11.8 24.4 22.1 38.4 30.7-27 23.6-61.7 36.7-97.7 36.7-82 0-148.5-66.5-148.5-148.5 0-39.4 15.6-77.1 43.5-105l79.1-79.1c27.8-27.8 65.6-43.5 105-43.5 82 0 148.5 66.5 148.5 148.5 0 17.9-3.2 35.5-9.4 51.9-4.6 12.4-18.5 18.7-30.9 14.1S317 292 321.6 279.6c4.2-11.1 6.4-23 6.4-35.1 0-55.5-45-100.5-100.5-100.5z\"\/><\/svg><\/span><\/a><h2 id=\"Section_Additional_Information\" class=\"header\">Additional Information<\/h2><\/div>\n\n<p><a href=\"https:\/\/otwo.com\/products\/automatic-transport-ventilators\/hand-held-automatic-ventilators-resuscitators\/carevent-ventilators\/\" rel=\"nofollow noopener noreferrer\">CAREven Official Page<\/a><\/p>\n\n<p><a href=\"\/Document\/CSUeFcPVO1ESPnVQ\/O-Two-CAREvent-Automatic-Transport-Ventilator-User-Manual.pdf\" class=\"docLink\"><span    class=\"fa-svg-icon docLink-icon\"><svg width=\"14\" height=\"14\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 384 512\" overflow=\"visible\"><!--! Font Awesome Pro 7.3.1 by @fontawesome - https:\/\/fontawesome.com License - https:\/\/fontawesome.com\/license (Commercial License) Copyright 2026 Fonticons, Inc. --><path d=\"M64 48l112 0 0 88c0 39.8 32.2 72 72 72l88 0 0 240c0 8.8-7.2 16-16 16L64 464c-8.8 0-16-7.2-16-16L48 64c0-8.8 7.2-16 16-16zM224 67.9l92.1 92.1-68.1 0c-13.3 0-24-10.7-24-24l0-68.1zM64 0C28.7 0 0 28.7 0 64L0 448c0 35.3 28.7 64 64 64l256 0c35.3 0 64-28.7 64-64l0-261.5c0-17-6.7-33.3-18.7-45.3L242.7 18.7C230.7 6.7 214.5 0 197.5 0L64 0zm56 256c-13.3 0-24 10.7-24 24s10.7 24 24 24l144 0c13.3 0 24-10.7 24-24s-10.7-24-24-24l-144 0zm0 96c-13.3 0-24 10.7-24 24s10.7 24 24 24l144 0c13.3 0 24-10.7 24-24s-10.7-24-24-24l-144 0z\"\/><\/svg><\/span>CAREvent User Manual<\/a><\/p>\n\n<p><a href=\"\/Document\/EkwePYRmOcHExBDx\/O-Two-CAREvent-ATV-Plus-MRI-Specification-Sheet.pdf\" class=\"docLink\"><span    class=\"fa-svg-icon docLink-icon\"><svg width=\"14\" height=\"14\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 384 512\" overflow=\"visible\"><!--! Font Awesome Pro 7.3.1 by @fontawesome - https:\/\/fontawesome.com License - https:\/\/fontawesome.com\/license (Commercial License) Copyright 2026 Fonticons, Inc. --><path d=\"M64 48l112 0 0 88c0 39.8 32.2 72 72 72l88 0 0 240c0 8.8-7.2 16-16 16L64 464c-8.8 0-16-7.2-16-16L48 64c0-8.8 7.2-16 16-16zM224 67.9l92.1 92.1-68.1 0c-13.3 0-24-10.7-24-24l0-68.1zM64 0C28.7 0 0 28.7 0 64L0 448c0 35.3 28.7 64 64 64l256 0c35.3 0 64-28.7 64-64l0-261.5c0-17-6.7-33.3-18.7-45.3L242.7 18.7C230.7 6.7 214.5 0 197.5 0L64 0zm56 256c-13.3 0-24 10.7-24 24s10.7 24 24 24l144 0c13.3 0 24-10.7 24-24s-10.7-24-24-24l-144 0zm0 96c-13.3 0-24 10.7-24 24s10.7 24 24 24l144 0c13.3 0 24-10.7 24-24s-10.7-24-24-24l-144 0z\"\/><\/svg><\/span>CAREvent Specification Sheet<\/a><\/p>","linked_wikis":[],"can_edit":true,"available_langids":["en","de","es","fr","it","jp","ko","nl","pt","ru","tr","zh"],"ancestors":[{"dataType":"wiki","wikiid":415933,"langid":"en","title":"O-Two Ventilator","display_title":"O-Two Ventilator","namespace":"CATEGORY","summary":"","url":"https:\/\/www.ifixit.com\/Device\/O-Two_Ventilator","text":"Repair guides, manuals, and troubleshooting help for O-Two Ventilator. Get everything you need to repair your O-Two Ventilator yourself.","image":{"id":2083144,"guid":"kcDi6sxuKOIGLDUr","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/kcDi6sxuKOIGLDUr.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/kcDi6sxuKOIGLDUr.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/kcDi6sxuKOIGLDUr.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/kcDi6sxuKOIGLDUr.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/kcDi6sxuKOIGLDUr.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/kcDi6sxuKOIGLDUr.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/kcDi6sxuKOIGLDUr.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/kcDi6sxuKOIGLDUr.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/kcDi6sxuKOIGLDUr.full"},"modified_date":1716920221,"author":1},{"dataType":"wiki","wikiid":414149,"langid":"en","title":"Ventilator","display_title":"Ventilator","namespace":"CATEGORY","summary":"Service and repair information for all makes and models of mechanical ventilation (assisted breathing) devices, including ICU, transport, neonatal, and PAP ventilators.","url":"https:\/\/www.ifixit.com\/Device\/Ventilator","text":"Service and repair information for all makes and models of mechanical ventilation (assisted breathing) devices, including ICU, transport, neonatal, and PAP ventilators.","image":{"id":3083122,"guid":"EnjtlsLZvrCsgvKZ","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EnjtlsLZvrCsgvKZ.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EnjtlsLZvrCsgvKZ.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EnjtlsLZvrCsgvKZ.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EnjtlsLZvrCsgvKZ.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EnjtlsLZvrCsgvKZ.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EnjtlsLZvrCsgvKZ.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EnjtlsLZvrCsgvKZ.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EnjtlsLZvrCsgvKZ.large","huge":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EnjtlsLZvrCsgvKZ.huge","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EnjtlsLZvrCsgvKZ.full"},"modified_date":1746025304,"author":668072},{"dataType":"wiki","wikiid":628736,"langid":"en","title":"Respiratory therapy","display_title":"Respiratory therapy","namespace":"CATEGORY","summary":"This folder contains technical information and documentation related to key devices used in respiratory therapy, essential for supporting and managing patients with respiratory conditions or ventilatory compromise. These devices are critical in hospital settings, outpatient clinics, and home care environments, and are typically operated by respiratory therapists, clinicians, and trained healthcare personnel.","url":"https:\/\/www.ifixit.com\/Device\/Respiratory_therapy","text":"This folder contains technical information and documentation related to key devices used in respiratory therapy, essential for supporting and managing patients with respiratory conditions or ventilatory compromise. These devices are critical in hospital settings, outpatient clinics, and home care environments, and are typically operated by respiratory therapists, clinicians, and trained healthcare personnel.","image":{"id":3638456,"guid":"LHyNeUDGJqEARZ14","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/LHyNeUDGJqEARZ14.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/LHyNeUDGJqEARZ14.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/LHyNeUDGJqEARZ14.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/LHyNeUDGJqEARZ14.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/LHyNeUDGJqEARZ14.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/LHyNeUDGJqEARZ14.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/LHyNeUDGJqEARZ14.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/LHyNeUDGJqEARZ14.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/LHyNeUDGJqEARZ14.full"},"modified_date":1792696876,"author":4626300},{"dataType":"wiki","wikiid":424438,"langid":"en","title":"Clinical Equipment","display_title":"Clinical Equipment","namespace":"CATEGORY","summary":"Service and repair information for medical devices most commonly used in a clinical setting.","url":"https:\/\/www.ifixit.com\/Device\/Clinical_Equipment","text":"Service and repair information for medical devices most commonly used in a clinical setting.","image":{"id":3076694,"guid":"q5qPhLJwOZejRdGi","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.large","huge":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.huge","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.full"},"modified_date":1792696444,"author":668072},{"dataType":"wiki","wikiid":17171,"langid":"en","title":"Medical Device","display_title":"Medical Device","namespace":"CATEGORY","summary":"Repair manuals and support documentation for devices used in a health care setting or procedure.","url":"https:\/\/www.ifixit.com\/Device\/Medical_Device","text":"Repair manuals and support documentation for devices used in a health care setting or procedure.","image":{"id":3076857,"guid":"yy4PBEHTeUqChV3B","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.large","huge":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.huge","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.full"},"modified_date":1792696418,"author":524640},{"dataType":"wiki","wikiid":11000,"langid":"en","title":"Root","display_title":"Root","namespace":"CATEGORY","summary":"","url":"https:\/\/www.ifixit.com\/Device\/Root","text":"Repair guides, manuals, and troubleshooting help for Root. Get everything you need to repair your Root yourself.","image":{"id":243440,"guid":"bhofAWyrmGwRFlak","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.full"},"modified_date":1705097689,"author":1}],"created_date":1597769721,"modified_date":1792520548,"is_troubleshooting":false,"authors":[{"userid":2166354,"percent":10,"username":"Justin Ogihara","avatar_url":"https:\/\/guide-images.cdn.ifixit.com\/igi\/qboLSPsUbuWKFV1K.thumbnail","profile_url":"https:\/\/www.ifixit.com\/User\/2166354\/Justin+Ogihara"},{"userid":3672148,"percent":45,"username":"Samson Wong","avatar_url":"https:\/\/assets.cdn.ifixit.com\/static\/images\/avatars\/User\/ifixit\/avatar-12.thumbnail","profile_url":"https:\/\/www.ifixit.com\/User\/3672148\/Samson+Wong"},{"userid":2989458,"percent":25,"username":"Austin Blakely","avatar_url":"https:\/\/guide-images.cdn.ifixit.com\/igi\/U2yM4NVNHFvF3FV3.thumbnail","profile_url":"https:\/\/www.ifixit.com\/User\/2989458\/Austin+Blakely"},{"userid":1,"percent":10,"username":"iRobot","avatar_url":"https:\/\/guide-images.cdn.ifixit.com\/igi\/AjiayMD6LWqf5FAL.thumbnail","profile_url":"https:\/\/www.ifixit.com\/User\/1\/iRobot"},{"userid":2,"percent":10,"username":"Kyle Wiens","avatar_url":"https:\/\/guide-images.cdn.ifixit.com\/igi\/MrphvTGGRcKHXRV2.thumbnail","profile_url":"https:\/\/www.ifixit.com\/User\/2\/Kyle+Wiens"}],"tags":[],"unreviewed_machine_translation":false,"flags":{"WIKI_BIOMED":{"flagid":"WIKI_BIOMED","title":"Medical Equipment","text":"Neither iFixit nor this repair information is endorsed by or affiliated with the medical device manufacturer.","isFontAwesome":true,"image":"notes-medical"}},"image":{"id":2260468,"guid":"FkeQrBGFKfluLwjy","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/FkeQrBGFKfluLwjy.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/FkeQrBGFKfluLwjy.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/FkeQrBGFKfluLwjy.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/FkeQrBGFKfluLwjy.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/FkeQrBGFKfluLwjy.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/FkeQrBGFKfluLwjy.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/FkeQrBGFKfluLwjy.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/FkeQrBGFKfluLwjy.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/FkeQrBGFKfluLwjy.full"},"documents":[{"documentid":24644,"guid":"I5hQDElFw4NV4NK3","date":1597769804,"size":165159,"pages":15,"filename":"O-Two-CareVent-User-manual.pdf","title":"O-Two-CareVent-User-manual.pdf","document_extension":"pdf","image":{"id":2182829,"guid":"1krDEjmxTl1GKoyB","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/1krDEjmxTl1GKoyB.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/1krDEjmxTl1GKoyB.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/1krDEjmxTl1GKoyB.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/1krDEjmxTl1GKoyB.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/1krDEjmxTl1GKoyB.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/1krDEjmxTl1GKoyB.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/1krDEjmxTl1GKoyB.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/1krDEjmxTl1GKoyB.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/1krDEjmxTl1GKoyB.full"},"url":"\/Document\/I5hQDElFw4NV4NK3\/O-Two-CareVent-User-manual.pdf","download_url":"https:\/\/www.ifixit.com\/Document\/Download\/I5hQDElFw4NV4NK3","display":"list"},{"documentid":2572276,"guid":"CSUeFcPVO1ESPnVQ","date":1791610213,"size":883134,"pages":16,"filename":"O-Two-CAREvent-Automatic-Transport-Ventilator-User-Manual.pdf","title":"CAREvent User Manual","document_extension":"pdf","image":{"id":6975804,"guid":"mohGMRsPDhWfRFpT","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/mohGMRsPDhWfRFpT.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/mohGMRsPDhWfRFpT.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/mohGMRsPDhWfRFpT.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/mohGMRsPDhWfRFpT.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/mohGMRsPDhWfRFpT.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/mohGMRsPDhWfRFpT.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/mohGMRsPDhWfRFpT.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/mohGMRsPDhWfRFpT.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/mohGMRsPDhWfRFpT.full"},"url":"\/Document\/CSUeFcPVO1ESPnVQ\/O-Two-CAREvent-Automatic-Transport-Ventilator-User-Manual.pdf","download_url":"https:\/\/www.ifixit.com\/Document\/Download\/CSUeFcPVO1ESPnVQ","display":null},{"documentid":2572999,"guid":"EkwePYRmOcHExBDx","date":1791623685,"size":1327602,"pages":3,"filename":"O-Two-CAREvent-ATV-Plus-MRI-Specification-Sheet.pdf","title":"CAREvent Specification Sheet","document_extension":"pdf","image":{"id":6976712,"guid":"EdJcOw54sMDZqCtB","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EdJcOw54sMDZqCtB.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EdJcOw54sMDZqCtB.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EdJcOw54sMDZqCtB.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EdJcOw54sMDZqCtB.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EdJcOw54sMDZqCtB.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EdJcOw54sMDZqCtB.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EdJcOw54sMDZqCtB.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EdJcOw54sMDZqCtB.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/EdJcOw54sMDZqCtB.full"},"url":"\/Document\/EkwePYRmOcHExBDx\/O-Two-CAREvent-ATV-Plus-MRI-Specification-Sheet.pdf","download_url":"https:\/\/www.ifixit.com\/Document\/Download\/EkwePYRmOcHExBDx","display":null}],"author":2166354,"display_title":"O-Two CAREvent","page_title":"O-Two CAREvent Repair","description":"Repair and service information for the O-Two CAREvent medical ventilator, used primarily in ATV+ and MRI settings.","children":[],"category_lists":{},"featured_guides":[],"guides":[],"related_wikis":[],"solutions_url":"https:\/\/www.ifixit.com\/Answers\/Device\/O-Two_CAREvent","repairability_score":null,"category_info":[],"source_revisionid":null,"diagrams":[],"info":[{"name":"condition_belt_wear","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_brake_pad_thickness_mm","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_controller_runtime_hours","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_coolant_ph","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_filter_pressure_drop","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_oil_life","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_oil_life_percent","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_tire_tread_depth_mm","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_water_hardness","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_battery_cycle_count","value":"on","inheritedFrom":"Clinical Equipment"},{"name":"condition_battery_health_percent","value":"on","inheritedFrom":"Clinical Equipment"},{"name":"condition_filter_life_percent","value":"on","inheritedFrom":"Respiratory therapy"}],"aliases":[],"maintenance_schedule":{"schedules":[],"notSuggested":[],"inheritedFrom":null,"langid":"en"}}