{"wikiid":419827,"langid":"en","namespace":"CATEGORY","title":"Drager Primus","revisionid":2625810,"contents_raw":"== ***Device Identification:*** ==\n\nThe Dr\u00e4ger Primus is an anesthesia workstation designated by model number SW 4.5n. The device includes a full suite of tools and devices to ventilate, monitor, and provide anesthesia to patients. The workstation is housed on a mobile rack which can be wheeled about as needed, and has room for various attachments and for gas tanks on the rack.\n\nThe Primus uses a the in-house user interface that is shared among other Drager platforms. This allows for a short and familiar learning process, despite being equipped with newer technology. It also means that technicians and experts trained in repairing the older devices can translate the same skills into repairing and operating the Primus.\n\nThis device is equipped with the E-Vent plus ventilator.\n\n== ***Technical Specifications:*** ==\n\n=== ***Device:*** ===\n\n* Weight \u2013 115kg (without vaporizers and cylinders)\n* Dimensions \u2013 137 cm x 80 cm x 80 cm\n* Device versions - Trolley version, wall version & ceiling version\n* Operating voltage \u2013 100-240 V, 45-65 Hz\n\n=== ***Function***: ===\n\n* Ventilator technology \u2013 Electronically-controlled E-Vent plus\n* APL-technology \u2013 Drager APL-Valve\n* Display screen \u2013 User-configurable 12.1\u201d TFT color\n* Vaporizers \u2013 2\n* Patient Gas Modules \u2013 in\/ex O2, N2O, CO2, Agent\n* Operating Modes \u2013 Manual, spontaneous, Volume Mode (IPPV), Pressure Mode (PCV)\n* Trigger \u2013 0.3-15 L\/min\n* Breathing Frequency \u2013 3-80 1\/min\n* Serial interfaces \u2013 2x RS232\n\n== ***Additional Information and Resources:*** ==\n\n* [https:\/\/www.draeger.com\/Products\/Content\/primus-pi-9069079-en-gb.pdf|Drager Products tab provides general information and functions of the device. Primus\u00ae Anaesthesia Workstations. (n.d.). L\u00fcbeck, Germany. Retrieved \u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0 November 10, 2020]\n* [https:\/\/www.draeger.com\/Library\/Content\/IfU_Primus_IE_SW_4.5n_EN_9053434.pdf|Instructions for use Drager Primus [Manual]. (n.d.). L\u00fcbeck, Germany. Retrieved \u00a0\u00a0\u00a0\u00a0\u00a0\u00a0 November 10, 2020]\n** The Drager Contents Library explains the use of the device\u2019s modes and settings, the set-up procedures, along with info on required maintenance and other technical data.\n* [https:\/\/www.draeger.com\/en_corp\/Corporate\/Coronavirus-COVID-19#anesthesia|Drager Corporate updates the operators with up to date Covid-19 related information on getting the most efficient use from their devices]\n* [https:\/\/pubs.asahq.org\/anesthesiology\/article\/106\/2\/289\/8059\/Accelerating-the-Washout-of-Inhalational|Crawford, M., Prinzhausen, H., & Petroz, G. (2007, February 01). Accelerating the Washout of Inhalational Anesthetics from the Dr\u00e4ger Primus Anesthetic Workstation Effect of Exchangeable Internal Components. Retrieved November 21, 2020]\n** An extensive study with detailed results on the effects and benefits of the device,  done in Toronto, partnered with Occupational Health and Safety Services, The Hospital for Sick Children.\n\nThe listed sources of information were mostly ++manufacturer++ provided information. Extensive data and information had been released with this device from the factory to ensure a resourceful database is available to users.\n\n[document|36670]\n\n[document|316712|Primus\/Primus Infinity Empowered Accessories Catalogue]","contents_json":{"type":"doc","content":[{"type":"paragraph","content":[]},{"type":"paragraph","content":[]},{"type":"heading","attrs":{"level":2},"content":[{"type":"text","text":"Device Identification:","marks":[{"type":"bold"}]}]},{"type":"paragraph","content":[{"type":"text","text":"The Dr\u00e4ger Primus is an anesthesia workstation designated by model number SW 4.5n. The device includes a full suite of tools and devices to ventilate, monitor, and provide anesthesia to patients. The workstation is housed on a mobile rack which can be wheeled about as needed, and has room for various attachments and for gas tanks on the rack."}]},{"type":"paragraph","content":[{"type":"text","text":"The Primus uses a the in-house user interface that is shared among other Drager platforms. This allows for a short and familiar learning process, despite being equipped with newer technology. It also means that technicians and experts trained in repairing the older devices can translate the same skills into repairing and operating the Primus."}]},{"type":"paragraph","content":[{"type":"text","text":"This device is equipped with the E-Vent plus ventilator."}]},{"type":"heading","attrs":{"level":2},"content":[{"type":"text","text":"Technical Specifications:","marks":[{"type":"bold"}]}]},{"type":"heading","attrs":{"level":3},"content":[{"type":"text","text":"Device:","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Weight \u2013 115kg (without vaporizers and cylinders)"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Dimensions \u2013 137 cm x 80 cm x 80 cm"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Device versions - Trolley version, wall version & ceiling version"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Operating voltage \u2013 100-240 V, 45-65 Hz"}]}]}]},{"type":"heading","attrs":{"level":3},"content":[{"type":"text","text":"Function","marks":[{"type":"bold"}]},{"type":"text","text":":"}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Ventilator technology \u2013 Electronically-controlled E-Vent plus"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"APL-technology \u2013 Drager APL-Valve"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Display screen \u2013 User-configurable 12.1\u201d TFT color"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Vaporizers \u2013 2"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Patient Gas Modules \u2013 in\/ex O2, N2O, CO2, Agent"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Operating Modes \u2013 Manual, spontaneous, Volume Mode (IPPV), Pressure Mode (PCV)"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Trigger \u2013 0.3-15 L\/min"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Breathing Frequency \u2013 3-80 1\/min"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"Serial interfaces \u2013 2x RS232"}]}]}]},{"type":"heading","attrs":{"level":2},"content":[{"type":"text","text":"Additional Information and Resources:","marks":[{"type":"bold"}]}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","marks":[{"type":"link","attrs":{"href":"https:\/\/www.draeger.com\/Products\/Content\/primus-pi-9069079-en-gb.pdf","target":false}}],"text":"Drager Products tab provides general information and functions of the device. Primus\u00ae Anaesthesia Workstations. (n.d.). L\u00fcbeck, Germany. Retrieved \u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0 November 10, 2020"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","marks":[{"type":"link","attrs":{"href":"https:\/\/www.draeger.com\/Library\/Content\/IfU_Primus_IE_SW_4.5n_EN_9053434.pdf","target":false}}],"text":"Instructions for use Drager Primus [Manual"},{"type":"text","text":". (n.d.). L\u00fcbeck, Germany. Retrieved \u00a0\u00a0\u00a0\u00a0\u00a0\u00a0 November 10, 2020]"}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"The Drager Contents Library explains the use of the device\u2019s modes and settings, the set-up procedures, along with info on required maintenance and other technical data."}]}]}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","marks":[{"type":"link","attrs":{"href":"https:\/\/www.draeger.com\/en_corp\/Corporate\/Coronavirus-COVID-19#anesthesia","target":false}}],"text":"Drager Corporate updates the operators with up to date Covid-19 related information on getting the most efficient use from their devices"}]}]},{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","marks":[{"type":"link","attrs":{"href":"https:\/\/pubs.asahq.org\/anesthesiology\/article\/106\/2\/289\/8059\/Accelerating-the-Washout-of-Inhalational","target":false}}],"text":"Crawford, M., Prinzhausen, H., & Petroz, G. (2007, February 01). Accelerating the Washout of Inhalational Anesthetics from the Dr\u00e4ger Primus Anesthetic Workstation Effect of Exchangeable Internal Components. Retrieved November 21, 2020"}]},{"type":"bulletList","attrs":[],"content":[{"type":"listItem","content":[{"type":"paragraph","content":[{"type":"text","text":"An extensive study with detailed results on the effects and benefits of the device,  done in Toronto, partnered with Occupational Health and Safety Services, The Hospital for Sick Children."}]}]}]}]}]},{"type":"paragraph","content":[{"type":"text","text":"The listed sources of information were mostly "},{"type":"text","text":"manufacturer","marks":[{"type":"underline"}]},{"type":"text","text":" provided information. Extensive data and information had been released with this device from the factory to ensure a resourceful database is available to users."}]},{"type":"paragraph","content":[{"type":"text","text":"Primus Anaesthetic Workstation Rev. 6.0 Technical Documentation","marks":[{"type":"siteLink","attrs":{"kind":"document","id":"36670","options":"","defaultText":"Primus Anaesthetic Workstation Rev. 6.0 Technical Documentation","newWindow":false,"href":"\/Document\/NxJFQ2dmGQ6UQ3KM\/Drager_Primus_Anaesthetic_Workstation_Rev._6.0_Technical_Documentation.pdf.pdf"}}]}]},{"type":"paragraph","content":[{"type":"text","text":"Primus\/Primus Infinity Empowered Accessories Catalogue","marks":[{"type":"siteLink","attrs":{"kind":"document","id":"316712","options":"","defaultText":null,"newWindow":false,"href":"\/Document\/eioPI4AMV3lPqeHW\/Primus-Primus-Infinity-Empowered-Accessories-Catalogue.pdf"}}]}]}]},"contents_rendered":"<div class=\"clearer\"><\/div><div class=\"headerContainer\"><a class=\"selfLink\" href=\"#Section__Device_Identification\" title=\"Link to this section\" aria-hidden=\"true\" tabindex=\"-1\"><span    class=\"fa-svg-icon\"><svg width=\"16\" height=\"16\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 576 512\" overflow=\"visible\"><!--! Font Awesome Pro 7.3.1 by @fontawesome - https:\/\/fontawesome.com License - https:\/\/fontawesome.com\/license (Commercial License) Copyright 2026 Fonticons, Inc. --><path d=\"M427.5 88c-21.4 0-42.2 6.9-59.3 19.4-11.4-11.7-24.4-22.1-38.4-30.7 27-23.6 61.7-36.7 97.7-36.7 82 0 148.5 66.5 148.5 148.5 0 39.4-15.6 77.1-43.5 105l-79.1 79.1c-27.8 27.8-65.6 43.5-105 43.5-82 0-148.5-66.5-148.5-148.5 0-17.9 3.2-35.5 9.4-51.9 4.6-12.4 18.5-18.7 30.9-14.1S259 220 254.4 232.4c-4.2 11.1-6.4 23-6.4 35.2 0 55.5 45 100.5 100.5 100.5 26.6 0 52.2-10.6 71-29.4l79.1-79.1C517.4 240.6 528 215.1 528 188.5 528 133 483 88 427.5 88zm-200 56c-26.6 0-52.2 10.6-71 29.4L77.4 252.5c-18.8 18.8-29.4 44.4-29.4 71 0 55.5 45 100.5 100.5 100.5 21.4 0 42.2-6.9 59.3-19.4 11.4 11.8 24.4 22.1 38.4 30.7-27 23.6-61.7 36.7-97.7 36.7-82 0-148.5-66.5-148.5-148.5 0-39.4 15.6-77.1 43.5-105l79.1-79.1c27.8-27.8 65.6-43.5 105-43.5 82 0 148.5 66.5 148.5 148.5 0 17.9-3.2 35.5-9.4 51.9-4.6 12.4-18.5 18.7-30.9 14.1S317 292 321.6 279.6c4.2-11.1 6.4-23 6.4-35.1 0-55.5-45-100.5-100.5-100.5z\"\/><\/svg><\/span><\/a><h2 id=\"Section__Device_Identification\" class=\"header\"><strong>Device Identification:<\/strong><\/h2><\/div>\n\n<p>The Dr\u00e4ger Primus is an anesthesia workstation designated by model number SW 4.5n. The device includes a full suite of tools and devices to ventilate, monitor, and provide anesthesia to patients. The workstation is housed on a mobile rack which can be wheeled about as needed, and has room for various attachments and for gas tanks on the rack.<\/p>\n\n<p>The Primus uses a the in-house user interface that is shared among other Drager platforms. This allows for a short and familiar learning process, despite being equipped with newer technology. It also means that technicians and experts trained in repairing the older devices can translate the same skills into repairing and operating the Primus.<\/p>\n\n<p>This device is equipped with the E-Vent plus ventilator.<\/p>\n\n<div class=\"clearer\"><\/div><div class=\"headerContainer\"><a class=\"selfLink\" href=\"#Section__Technical_Specifications\" title=\"Link to this section\" aria-hidden=\"true\" tabindex=\"-1\"><span    class=\"fa-svg-icon\"><svg width=\"16\" height=\"16\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 576 512\" overflow=\"visible\"><!--! Font Awesome Pro 7.3.1 by @fontawesome - https:\/\/fontawesome.com License - https:\/\/fontawesome.com\/license (Commercial License) Copyright 2026 Fonticons, Inc. --><path d=\"M427.5 88c-21.4 0-42.2 6.9-59.3 19.4-11.4-11.7-24.4-22.1-38.4-30.7 27-23.6 61.7-36.7 97.7-36.7 82 0 148.5 66.5 148.5 148.5 0 39.4-15.6 77.1-43.5 105l-79.1 79.1c-27.8 27.8-65.6 43.5-105 43.5-82 0-148.5-66.5-148.5-148.5 0-17.9 3.2-35.5 9.4-51.9 4.6-12.4 18.5-18.7 30.9-14.1S259 220 254.4 232.4c-4.2 11.1-6.4 23-6.4 35.2 0 55.5 45 100.5 100.5 100.5 26.6 0 52.2-10.6 71-29.4l79.1-79.1C517.4 240.6 528 215.1 528 188.5 528 133 483 88 427.5 88zm-200 56c-26.6 0-52.2 10.6-71 29.4L77.4 252.5c-18.8 18.8-29.4 44.4-29.4 71 0 55.5 45 100.5 100.5 100.5 21.4 0 42.2-6.9 59.3-19.4 11.4 11.8 24.4 22.1 38.4 30.7-27 23.6-61.7 36.7-97.7 36.7-82 0-148.5-66.5-148.5-148.5 0-39.4 15.6-77.1 43.5-105l79.1-79.1c27.8-27.8 65.6-43.5 105-43.5 82 0 148.5 66.5 148.5 148.5 0 17.9-3.2 35.5-9.4 51.9-4.6 12.4-18.5 18.7-30.9 14.1S317 292 321.6 279.6c4.2-11.1 6.4-23 6.4-35.1 0-55.5-45-100.5-100.5-100.5z\"\/><\/svg><\/span><\/a><h2 id=\"Section__Technical_Specifications\" class=\"header\"><strong>Technical Specifications:<\/strong><\/h2><\/div>\n\n<div class=\"clearer\"><\/div><div class=\"headerContainer\"><a class=\"selfLink\" href=\"#Section__Device\" title=\"Link to this section\" aria-hidden=\"true\" tabindex=\"-1\"><span    class=\"fa-svg-icon\"><svg width=\"16\" height=\"16\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 576 512\" overflow=\"visible\"><!--! Font Awesome Pro 7.3.1 by @fontawesome - https:\/\/fontawesome.com License - https:\/\/fontawesome.com\/license (Commercial License) Copyright 2026 Fonticons, Inc. --><path d=\"M427.5 88c-21.4 0-42.2 6.9-59.3 19.4-11.4-11.7-24.4-22.1-38.4-30.7 27-23.6 61.7-36.7 97.7-36.7 82 0 148.5 66.5 148.5 148.5 0 39.4-15.6 77.1-43.5 105l-79.1 79.1c-27.8 27.8-65.6 43.5-105 43.5-82 0-148.5-66.5-148.5-148.5 0-17.9 3.2-35.5 9.4-51.9 4.6-12.4 18.5-18.7 30.9-14.1S259 220 254.4 232.4c-4.2 11.1-6.4 23-6.4 35.2 0 55.5 45 100.5 100.5 100.5 26.6 0 52.2-10.6 71-29.4l79.1-79.1C517.4 240.6 528 215.1 528 188.5 528 133 483 88 427.5 88zm-200 56c-26.6 0-52.2 10.6-71 29.4L77.4 252.5c-18.8 18.8-29.4 44.4-29.4 71 0 55.5 45 100.5 100.5 100.5 21.4 0 42.2-6.9 59.3-19.4 11.4 11.8 24.4 22.1 38.4 30.7-27 23.6-61.7 36.7-97.7 36.7-82 0-148.5-66.5-148.5-148.5 0-39.4 15.6-77.1 43.5-105l79.1-79.1c27.8-27.8 65.6-43.5 105-43.5 82 0 148.5 66.5 148.5 148.5 0 17.9-3.2 35.5-9.4 51.9-4.6 12.4-18.5 18.7-30.9 14.1S317 292 321.6 279.6c4.2-11.1 6.4-23 6.4-35.1 0-55.5-45-100.5-100.5-100.5z\"\/><\/svg><\/span><\/a><h3 id=\"Section__Device\" class=\"header\"><strong>Device:<\/strong><\/h3><\/div>\n\n<ul><li>Weight \u2013 115kg (without vaporizers and cylinders)<\/li><li>Dimensions \u2013 137 cm x 80 cm x 80 cm<\/li><li>Device versions - Trolley version, wall version &amp; ceiling version<\/li><li>Operating voltage \u2013 100-240 V, 45-65 Hz<\/li><\/ul>\n\n<div class=\"clearer\"><\/div><div class=\"headerContainer\"><a class=\"selfLink\" href=\"#Section__Function\" title=\"Link to this section\" aria-hidden=\"true\" tabindex=\"-1\"><span    class=\"fa-svg-icon\"><svg width=\"16\" height=\"16\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 576 512\" overflow=\"visible\"><!--! Font Awesome Pro 7.3.1 by @fontawesome - https:\/\/fontawesome.com License - https:\/\/fontawesome.com\/license (Commercial License) Copyright 2026 Fonticons, Inc. --><path d=\"M427.5 88c-21.4 0-42.2 6.9-59.3 19.4-11.4-11.7-24.4-22.1-38.4-30.7 27-23.6 61.7-36.7 97.7-36.7 82 0 148.5 66.5 148.5 148.5 0 39.4-15.6 77.1-43.5 105l-79.1 79.1c-27.8 27.8-65.6 43.5-105 43.5-82 0-148.5-66.5-148.5-148.5 0-17.9 3.2-35.5 9.4-51.9 4.6-12.4 18.5-18.7 30.9-14.1S259 220 254.4 232.4c-4.2 11.1-6.4 23-6.4 35.2 0 55.5 45 100.5 100.5 100.5 26.6 0 52.2-10.6 71-29.4l79.1-79.1C517.4 240.6 528 215.1 528 188.5 528 133 483 88 427.5 88zm-200 56c-26.6 0-52.2 10.6-71 29.4L77.4 252.5c-18.8 18.8-29.4 44.4-29.4 71 0 55.5 45 100.5 100.5 100.5 21.4 0 42.2-6.9 59.3-19.4 11.4 11.8 24.4 22.1 38.4 30.7-27 23.6-61.7 36.7-97.7 36.7-82 0-148.5-66.5-148.5-148.5 0-39.4 15.6-77.1 43.5-105l79.1-79.1c27.8-27.8 65.6-43.5 105-43.5 82 0 148.5 66.5 148.5 148.5 0 17.9-3.2 35.5-9.4 51.9-4.6 12.4-18.5 18.7-30.9 14.1S317 292 321.6 279.6c4.2-11.1 6.4-23 6.4-35.1 0-55.5-45-100.5-100.5-100.5z\"\/><\/svg><\/span><\/a><h3 id=\"Section__Function\" class=\"header\"><strong>Function<\/strong>:<\/h3><\/div>\n\n<ul><li>Ventilator technology \u2013 Electronically-controlled E-Vent plus<\/li><li>APL-technology \u2013 Drager APL-Valve<\/li><li>Display screen \u2013 User-configurable 12.1\u201d TFT color<\/li><li>Vaporizers \u2013 2<\/li><li>Patient Gas Modules \u2013 in\/ex O2, N2O, CO2, Agent<\/li><li>Operating Modes \u2013 Manual, spontaneous, Volume Mode (IPPV), Pressure Mode (PCV)<\/li><li>Trigger \u2013 0.3-15 L\/min<\/li><li>Breathing Frequency \u2013 3-80 1\/min<\/li><li>Serial interfaces \u2013 2x RS232<\/li><\/ul>\n\n<div class=\"clearer\"><\/div><div class=\"headerContainer\"><a class=\"selfLink\" href=\"#Section__Additional_Information_and_Resources\" title=\"Link to this section\" aria-hidden=\"true\" tabindex=\"-1\"><span    class=\"fa-svg-icon\"><svg width=\"16\" height=\"16\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 576 512\" overflow=\"visible\"><!--! Font Awesome Pro 7.3.1 by @fontawesome - https:\/\/fontawesome.com License - https:\/\/fontawesome.com\/license (Commercial License) Copyright 2026 Fonticons, Inc. --><path d=\"M427.5 88c-21.4 0-42.2 6.9-59.3 19.4-11.4-11.7-24.4-22.1-38.4-30.7 27-23.6 61.7-36.7 97.7-36.7 82 0 148.5 66.5 148.5 148.5 0 39.4-15.6 77.1-43.5 105l-79.1 79.1c-27.8 27.8-65.6 43.5-105 43.5-82 0-148.5-66.5-148.5-148.5 0-17.9 3.2-35.5 9.4-51.9 4.6-12.4 18.5-18.7 30.9-14.1S259 220 254.4 232.4c-4.2 11.1-6.4 23-6.4 35.2 0 55.5 45 100.5 100.5 100.5 26.6 0 52.2-10.6 71-29.4l79.1-79.1C517.4 240.6 528 215.1 528 188.5 528 133 483 88 427.5 88zm-200 56c-26.6 0-52.2 10.6-71 29.4L77.4 252.5c-18.8 18.8-29.4 44.4-29.4 71 0 55.5 45 100.5 100.5 100.5 21.4 0 42.2-6.9 59.3-19.4 11.4 11.8 24.4 22.1 38.4 30.7-27 23.6-61.7 36.7-97.7 36.7-82 0-148.5-66.5-148.5-148.5 0-39.4 15.6-77.1 43.5-105l79.1-79.1c27.8-27.8 65.6-43.5 105-43.5 82 0 148.5 66.5 148.5 148.5 0 17.9-3.2 35.5-9.4 51.9-4.6 12.4-18.5 18.7-30.9 14.1S317 292 321.6 279.6c4.2-11.1 6.4-23 6.4-35.1 0-55.5-45-100.5-100.5-100.5z\"\/><\/svg><\/span><\/a><h2 id=\"Section__Additional_Information_and_Resources\" class=\"header\"><strong>Additional Information and Resources:<\/strong><\/h2><\/div>\n\n<ul><li><a href=\"https:\/\/www.draeger.com\/Products\/Content\/primus-pi-9069079-en-gb.pdf\" rel=\"nofollow noopener noreferrer\">Drager Products tab provides general information and functions of the device. Primus\u00ae Anaesthesia Workstations. (n.d.). L\u00fcbeck, Germany. Retrieved \u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0 November 10, 2020<\/a><\/li><li><a href=\"https:\/\/www.draeger.com\/Library\/Content\/IfU_Primus_IE_SW_4.5n_EN_9053434.pdf\" rel=\"nofollow noopener noreferrer\">Instructions for use Drager Primus [Manual<\/a>. (n.d.). L\u00fcbeck, Germany. Retrieved \u00a0\u00a0\u00a0\u00a0\u00a0\u00a0 November 10, 2020]<ul><li>The Drager Contents Library explains the use of the device\u2019s modes and settings, the set-up procedures, along with info on required maintenance and other technical data.<\/li><\/ul><\/li><li><a href=\"https:\/\/www.draeger.com\/en_corp\/Corporate\/Coronavirus-COVID-19#anesthesia\" rel=\"nofollow noopener noreferrer\">Drager Corporate updates the operators with up to date Covid-19 related information on getting the most efficient use from their devices<\/a><\/li><li><a href=\"https:\/\/pubs.asahq.org\/anesthesiology\/article\/106\/2\/289\/8059\/Accelerating-the-Washout-of-Inhalational\" rel=\"nofollow noopener noreferrer\">Crawford, M., Prinzhausen, H., &amp; Petroz, G. (2007, February 01). Accelerating the Washout of Inhalational Anesthetics from the Dr\u00e4ger Primus Anesthetic Workstation Effect of Exchangeable Internal Components. Retrieved November 21, 2020<\/a><ul><li>An extensive study with detailed results on the effects and benefits of the device,  done in Toronto, partnered with Occupational Health and Safety Services, The Hospital for Sick Children.<\/li><\/ul><\/li><\/ul>\n\n<p>The listed sources of information were mostly <ins>manufacturer<\/ins> provided information. Extensive data and information had been released with this device from the factory to ensure a resourceful database is available to users.<\/p>\n\n<p><a href=\"\/Document\/NxJFQ2dmGQ6UQ3KM\/Drager_Primus_Anaesthetic_Workstation_Rev._6.0_Technical_Documentation.pdf.pdf\" class=\"docLink\"><span    class=\"fa-svg-icon docLink-icon\"><svg width=\"14\" height=\"14\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 384 512\" overflow=\"visible\"><!--! Font Awesome Pro 7.3.1 by @fontawesome - https:\/\/fontawesome.com License - https:\/\/fontawesome.com\/license (Commercial License) Copyright 2026 Fonticons, Inc. --><path d=\"M64 48l112 0 0 88c0 39.8 32.2 72 72 72l88 0 0 240c0 8.8-7.2 16-16 16L64 464c-8.8 0-16-7.2-16-16L48 64c0-8.8 7.2-16 16-16zM224 67.9l92.1 92.1-68.1 0c-13.3 0-24-10.7-24-24l0-68.1zM64 0C28.7 0 0 28.7 0 64L0 448c0 35.3 28.7 64 64 64l256 0c35.3 0 64-28.7 64-64l0-261.5c0-17-6.7-33.3-18.7-45.3L242.7 18.7C230.7 6.7 214.5 0 197.5 0L64 0zm56 256c-13.3 0-24 10.7-24 24s10.7 24 24 24l144 0c13.3 0 24-10.7 24-24s-10.7-24-24-24l-144 0zm0 96c-13.3 0-24 10.7-24 24s10.7 24 24 24l144 0c13.3 0 24-10.7 24-24s-10.7-24-24-24l-144 0z\"\/><\/svg><\/span>Primus Anaesthetic Workstation Rev. 6.0 Technical Documentation<\/a><\/p>\n\n<p><a href=\"\/Document\/eioPI4AMV3lPqeHW\/Primus-Primus-Infinity-Empowered-Accessories-Catalogue.pdf\" class=\"docLink\"><span    class=\"fa-svg-icon docLink-icon\"><svg width=\"14\" height=\"14\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\" viewBox=\"0 0 384 512\" overflow=\"visible\"><!--! Font Awesome Pro 7.3.1 by @fontawesome - https:\/\/fontawesome.com License - https:\/\/fontawesome.com\/license (Commercial License) Copyright 2026 Fonticons, Inc. --><path d=\"M64 48l112 0 0 88c0 39.8 32.2 72 72 72l88 0 0 240c0 8.8-7.2 16-16 16L64 464c-8.8 0-16-7.2-16-16L48 64c0-8.8 7.2-16 16-16zM224 67.9l92.1 92.1-68.1 0c-13.3 0-24-10.7-24-24l0-68.1zM64 0C28.7 0 0 28.7 0 64L0 448c0 35.3 28.7 64 64 64l256 0c35.3 0 64-28.7 64-64l0-261.5c0-17-6.7-33.3-18.7-45.3L242.7 18.7C230.7 6.7 214.5 0 197.5 0L64 0zm56 256c-13.3 0-24 10.7-24 24s10.7 24 24 24l144 0c13.3 0 24-10.7 24-24s-10.7-24-24-24l-144 0zm0 96c-13.3 0-24 10.7-24 24s10.7 24 24 24l144 0c13.3 0 24-10.7 24-24s-10.7-24-24-24l-144 0z\"\/><\/svg><\/span>Primus\/Primus Infinity Empowered Accessories Catalogue<\/a><\/p>","linked_wikis":[],"can_edit":true,"available_langids":["en","de","es","fr","it","jp","ko","nl","pt","ru","tr","zh"],"ancestors":[{"dataType":"wiki","wikiid":415684,"langid":"en","title":"Drager Anesthesia","display_title":"Dr\u00e4ger Anesthesia","namespace":"CATEGORY","summary":"Support, service, and repair information for Drager Anesthesia machines.","url":"https:\/\/www.ifixit.com\/Device\/Drager_Anesthesia","text":"Support, service, and repair information for Drager Anesthesia machines.","image":{"id":2578689,"guid":"2AFHKbXRsI1YQgbR","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/2AFHKbXRsI1YQgbR.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/2AFHKbXRsI1YQgbR.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/2AFHKbXRsI1YQgbR.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/2AFHKbXRsI1YQgbR.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/2AFHKbXRsI1YQgbR.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/2AFHKbXRsI1YQgbR.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/2AFHKbXRsI1YQgbR.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/2AFHKbXRsI1YQgbR.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/2AFHKbXRsI1YQgbR.full"},"modified_date":1711400443,"author":1},{"dataType":"wiki","wikiid":416112,"langid":"en","title":"Anesthesia System","display_title":"Anesthesia System","namespace":"CATEGORY","summary":"Repair information and supporting documentation for anesthesia machines of all kinds.","url":"https:\/\/www.ifixit.com\/Device\/Anesthesia_System","text":"Repair information and supporting documentation for anesthesia machines of all kinds.","image":{"id":3076632,"guid":"xyNtGqr4CeXwA2g1","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/xyNtGqr4CeXwA2g1.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/xyNtGqr4CeXwA2g1.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/xyNtGqr4CeXwA2g1.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/xyNtGqr4CeXwA2g1.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/xyNtGqr4CeXwA2g1.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/xyNtGqr4CeXwA2g1.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/xyNtGqr4CeXwA2g1.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/xyNtGqr4CeXwA2g1.large","huge":"https:\/\/guide-images.cdn.ifixit.com\/igi\/xyNtGqr4CeXwA2g1.huge","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/xyNtGqr4CeXwA2g1.full"},"modified_date":1704819427,"author":668072},{"dataType":"wiki","wikiid":424438,"langid":"en","title":"Clinical Equipment","display_title":"Clinical Equipment","namespace":"CATEGORY","summary":"Service and repair information for medical devices most commonly used in a clinical setting.","url":"https:\/\/www.ifixit.com\/Device\/Clinical_Equipment","text":"Service and repair information for medical devices most commonly used in a clinical setting.","image":{"id":3076694,"guid":"q5qPhLJwOZejRdGi","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.large","huge":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.huge","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/q5qPhLJwOZejRdGi.full"},"modified_date":1792696444,"author":668072},{"dataType":"wiki","wikiid":17171,"langid":"en","title":"Medical Device","display_title":"Medical Device","namespace":"CATEGORY","summary":"Repair manuals and support documentation for devices used in a health care setting or procedure.","url":"https:\/\/www.ifixit.com\/Device\/Medical_Device","text":"Repair manuals and support documentation for devices used in a health care setting or procedure.","image":{"id":3076857,"guid":"yy4PBEHTeUqChV3B","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.large","huge":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.huge","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/yy4PBEHTeUqChV3B.full"},"modified_date":1792696418,"author":524640},{"dataType":"wiki","wikiid":11000,"langid":"en","title":"Root","display_title":"Root","namespace":"CATEGORY","summary":"","url":"https:\/\/www.ifixit.com\/Device\/Root","text":"Repair guides, manuals, and troubleshooting help for Root. Get everything you need to repair your Root yourself.","image":{"id":243440,"guid":"bhofAWyrmGwRFlak","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/bhofAWyrmGwRFlak.full"},"modified_date":1705097689,"author":1}],"created_date":1589762167,"modified_date":1792623188,"is_troubleshooting":false,"authors":[{"userid":1,"percent":46.67,"username":"iRobot","avatar_url":"https:\/\/guide-images.cdn.ifixit.com\/igi\/AjiayMD6LWqf5FAL.thumbnail","profile_url":"https:\/\/www.ifixit.com\/User\/1\/iRobot"},{"userid":3674009,"percent":26.67,"username":"Seyed-Mohammad Hosseiny","avatar_url":"https:\/\/guide-images.cdn.ifixit.com\/igi\/2IfHlBTYBRGP2DxR.thumbnail","profile_url":"https:\/\/www.ifixit.com\/User\/3674009\/Seyed-Mohammad+Hosseiny"},{"userid":2989458,"percent":13.33,"username":"Austin Blakely","avatar_url":"https:\/\/guide-images.cdn.ifixit.com\/igi\/U2yM4NVNHFvF3FV3.thumbnail","profile_url":"https:\/\/www.ifixit.com\/User\/2989458\/Austin+Blakely"},{"userid":668072,"percent":6.67,"username":"Jeff Suovanen","avatar_url":"https:\/\/guide-images.cdn.ifixit.com\/igi\/uYn4Ihpxm4PJCBrU.thumbnail","profile_url":"https:\/\/www.ifixit.com\/User\/668072\/Jeff+Suovanen"},{"userid":2012381,"percent":6.67,"username":"eCuration","avatar_url":"https:\/\/guide-images.cdn.ifixit.com\/igi\/vxmZa1BuxNTWWnGf.thumbnail","profile_url":"https:\/\/www.ifixit.com\/User\/2012381\/eCuration"}],"tags":[],"unreviewed_machine_translation":false,"flags":{"WIKI_BIOMED":{"flagid":"WIKI_BIOMED","title":"Medical Equipment","text":"Neither iFixit nor this repair information is endorsed by or affiliated with the medical device manufacturer.","isFontAwesome":true,"image":"notes-medical"}},"image":{"id":2261079,"guid":"r5bXvZxTZPOBZBXk","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/r5bXvZxTZPOBZBXk.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/r5bXvZxTZPOBZBXk.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/r5bXvZxTZPOBZBXk.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/r5bXvZxTZPOBZBXk.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/r5bXvZxTZPOBZBXk.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/r5bXvZxTZPOBZBXk.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/r5bXvZxTZPOBZBXk.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/r5bXvZxTZPOBZBXk.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/r5bXvZxTZPOBZBXk.full"},"documents":[{"documentid":14481,"guid":"pa43QAE4PGJmBXg6","date":1589764647,"size":3913722,"pages":100,"filename":"Drager Primus Anaesthetic Workstation Software 2n User manual.pdf","title":"Drager Primus Anaesthetic Workstation Software 2n User manual.pdf","document_extension":"pdf","image":{"id":2099942,"guid":"Z5OYFfrEQ4IYYDUM","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/Z5OYFfrEQ4IYYDUM.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/Z5OYFfrEQ4IYYDUM.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/Z5OYFfrEQ4IYYDUM.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/Z5OYFfrEQ4IYYDUM.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/Z5OYFfrEQ4IYYDUM.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/Z5OYFfrEQ4IYYDUM.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/Z5OYFfrEQ4IYYDUM.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/Z5OYFfrEQ4IYYDUM.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/Z5OYFfrEQ4IYYDUM.full"},"url":"\/Document\/pa43QAE4PGJmBXg6\/Drager%20Primus%20Anaesthetic%20Workstation%20Software%202n%20User%20manual.pdf","download_url":"https:\/\/www.ifixit.com\/Document\/Download\/pa43QAE4PGJmBXg6","display":"list"},{"documentid":14483,"guid":"TwJ63dv3PxJ2v6Nt","date":1589764647,"size":25619954,"pages":336,"filename":"Drager Primus Anaesthetic Workstation Rev. 6.0 Technical Documentation.pdf","title":"Drager Primus Anaesthetic Workstation Rev. 6.0 Technical Documentation.pdf","document_extension":"pdf","image":{"id":2099946,"guid":"HEOF6cN2KwYetPik","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/HEOF6cN2KwYetPik.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/HEOF6cN2KwYetPik.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/HEOF6cN2KwYetPik.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/HEOF6cN2KwYetPik.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/HEOF6cN2KwYetPik.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/HEOF6cN2KwYetPik.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/HEOF6cN2KwYetPik.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/HEOF6cN2KwYetPik.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/HEOF6cN2KwYetPik.full"},"url":"\/Document\/TwJ63dv3PxJ2v6Nt\/Drager%20Primus%20Anaesthetic%20Workstation%20Rev.%206.0%20Technical%20Documentation.pdf","download_url":"https:\/\/www.ifixit.com\/Document\/Download\/TwJ63dv3PxJ2v6Nt","display":"list"},{"documentid":36670,"guid":"NxJFQ2dmGQ6UQ3KM","date":1716605940,"size":25619954,"pages":336,"filename":"Drager_Primus_Anaesthetic_Workstation_Rev._6.0_Technical_Documentation.pdf.pdf","title":"Drager_Primus_Anaesthetic_Workstation_Rev._6.0_Technical_Documentation.pdf.pdf","document_extension":"pdf","image":{"id":3327195,"guid":"QPPQ4tPWQEXwOvi6","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/QPPQ4tPWQEXwOvi6.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/QPPQ4tPWQEXwOvi6.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/QPPQ4tPWQEXwOvi6.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/QPPQ4tPWQEXwOvi6.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/QPPQ4tPWQEXwOvi6.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/QPPQ4tPWQEXwOvi6.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/QPPQ4tPWQEXwOvi6.medium","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/QPPQ4tPWQEXwOvi6.full"},"url":"\/Document\/NxJFQ2dmGQ6UQ3KM\/Drager_Primus_Anaesthetic_Workstation_Rev._6.0_Technical_Documentation.pdf.pdf","download_url":"https:\/\/www.ifixit.com\/Document\/Download\/NxJFQ2dmGQ6UQ3KM","display":null},{"documentid":316712,"guid":"eioPI4AMV3lPqeHW","date":1783737692,"size":3010575,"pages":32,"filename":"Primus-Primus-Infinity-Empowered-Accessories-Catalogue.pdf","title":"Primus\/Primus Infinity Empowered Accessories Catalogue","document_extension":"pdf","image":{"id":4535554,"guid":"DaRMHNrL3XKfoVHf","mini":"https:\/\/guide-images.cdn.ifixit.com\/igi\/DaRMHNrL3XKfoVHf.mini","thumbnail":"https:\/\/guide-images.cdn.ifixit.com\/igi\/DaRMHNrL3XKfoVHf.thumbnail","140x105":"https:\/\/guide-images.cdn.ifixit.com\/igi\/DaRMHNrL3XKfoVHf.140x105","200x150":"https:\/\/guide-images.cdn.ifixit.com\/igi\/DaRMHNrL3XKfoVHf.200x150","standard":"https:\/\/guide-images.cdn.ifixit.com\/igi\/DaRMHNrL3XKfoVHf.standard","440x330":"https:\/\/guide-images.cdn.ifixit.com\/igi\/DaRMHNrL3XKfoVHf.440x330","medium":"https:\/\/guide-images.cdn.ifixit.com\/igi\/DaRMHNrL3XKfoVHf.medium","large":"https:\/\/guide-images.cdn.ifixit.com\/igi\/DaRMHNrL3XKfoVHf.large","original":"https:\/\/guide-images.cdn.ifixit.com\/igi\/DaRMHNrL3XKfoVHf.full"},"url":"\/Document\/eioPI4AMV3lPqeHW\/Primus-Primus-Infinity-Empowered-Accessories-Catalogue.pdf","download_url":"https:\/\/www.ifixit.com\/Document\/Download\/eioPI4AMV3lPqeHW","display":null}],"author":1,"display_title":"Dr\u00e4ger Primus","page_title":"Dr\u00e4ger Primus Repair","description":"Repair and service information for the Drager Primus, an anesthesia workstation SW 4.5n","children":[],"category_lists":{},"featured_guides":[],"guides":[],"related_wikis":[],"solutions_url":"https:\/\/www.ifixit.com\/Answers\/Device\/Drager_Primus","repairability_score":null,"category_info":[],"source_revisionid":null,"diagrams":[],"info":[{"name":"Device Brand","value":"Drager","inheritedFrom":null},{"name":"Device Type","value":"Anesthesia Machine","inheritedFrom":null},{"name":"model_numbers","value":"9120.300, 9025864","inheritedFrom":null},{"name":"condition_belt_wear","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_brake_pad_thickness_mm","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_controller_runtime_hours","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_coolant_ph","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_filter_life_percent","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_filter_pressure_drop","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_oil_life","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_oil_life_percent","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_tire_tread_depth_mm","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_water_hardness","value":"off","inheritedFrom":"Medical Device"},{"name":"condition_battery_cycle_count","value":"on","inheritedFrom":"Clinical Equipment"},{"name":"condition_battery_health_percent","value":"on","inheritedFrom":"Clinical Equipment"}],"aliases":[],"maintenance_schedule":{"schedules":[{"wiki_maintenance_scheduleid":526,"task":"Replace battery","status":"active","guideid":null,"parts":null,"triggers":[{"trigger_type":"time","metric":"elapsed_time","unit":"years","first_interval":null,"repeat_interval":2,"comparator":null,"threshold":null}],"inheritedFrom":"Anesthesia System","switchedBy":null,"repeatedBy":null,"rationale":null,"inheritedOn":true},{"wiki_maintenance_scheduleid":527,"task":"Replace O-rings","status":"active","guideid":null,"parts":null,"triggers":[{"trigger_type":"time","metric":"elapsed_time","unit":"years","first_interval":null,"repeat_interval":1,"comparator":null,"threshold":null}],"inheritedFrom":"Anesthesia System","switchedBy":null,"repeatedBy":null,"rationale":null,"inheritedOn":true}],"notSuggested":[],"inheritedFrom":"Anesthesia System","langid":"en"}}