Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement

Replace the battery in a 6000 series Welch Allyn Connex Vital Signs Monitor.

  • Author: kyle2
  • Difficulty: Easy

Battery replacement is recommended every 300 charge cycles. Use this guide to replace the battery in your 6000 series Welch Allyn Connex Vital Signs Monitor.

PLEASE NOTE: This repair guide was developed by the iFixit team based on Welch Allyn's own service manual. Neither iFixit nor this repair guide is endorsed by or affiliated with Welch Allyn.

Step 1 Battery  ¶ 

Place the monitor upside down on a clean surface. Insert a coin into the slot and push to open. Choose a coin that fits comfortably into the slot.

Step 1 Battery  ¶ 

  • Place the monitor upside down on a clean surface.

  • Insert a coin into the slot and push to open.

  • Choose a coin that fits comfortably into the slot.

  • Remove the battery cover.

Step 2  ¶ 

Use the plastic label to remove the battery from its recess. Use the plastic label to remove the battery from its recess. Use the plastic label to remove the battery from its recess.

Step 2  ¶ 

  • Use the plastic label to remove the battery from its recess.