Skip to main content

Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement

What you need

  1. Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement, Stand: step 1, image 1 of 2 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement, Stand: step 1, image 2 of 2
    • Unplug the AC power connection cable.

  2. Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 2, image 1 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 2, image 2 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 2, image 3 of 3
    • Place your thumb and forefinger on the blood pressure (NIBP) hose connector. Squeeze the side tabs until the connector releases.

    • Always grasp the hose by the connector. Do not pull on the hose itself.

    • Pull the connector away from the connector port.

  3. Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 3, image 1 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 3, image 2 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 3, image 3 of 3
    • Place your thumb and forefinger on the Pulse oximetry (SpO2 or combined SpO2/SpHb) cable connector. Squeeze the side tabs until the connector releases.

    • Always grasp the cable by the connector. Do not pull on the cable itself.

    • Pull the connector away from the connector port.

  4. Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 4, image 1 of 2 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 4, image 2 of 2
    • Grasp the temperature probe and pull it up to remove it from the monitor.

  5. Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 5, image 1 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 5, image 2 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 5, image 3 of 3
    • Remove the cover of the temperature module by pressing the tab and sliding the cover to the right.

  6. Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 6, image 1 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 6, image 2 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 6, image 3 of 3
    • Depress the spring tab on the temperature probe cable connector and withdraw it from the probe port.

  7. Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 7, image 1 of 1
    • Remove the flathead screw on the USB networking door.

    • Loosen the captive Phillips #2 screw securing the monitor to the stand.

  8. Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 8, image 1 of 1
    • Holding the monitor securely, open the USB networking door.

    • You may have to lift the monitor slightly to allow the door to swing out from the stand mount.

  9. Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 9, image 1 of 2 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 9, image 2 of 2
    • Detach any accessory USB cables from USB ports on the monitor.

  10. Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 10, image 1 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 10, image 2 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 10, image 3 of 3
    • Slide the monitor off of the stand tray.

  11. Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement, Battery: step 11, image 1 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement, Battery: step 11, image 2 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement, Battery: step 11, image 3 of 3
    • Place the monitor upside down on a clean surface.

    • Insert a coin into the slot and push to open.

    • Choose a coin that fits comfortably into the slot.

    • Remove the battery cover.

  12. Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 12, image 1 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 12, image 2 of 3 Welch Allyn Connex Vital Signs Monitor 6000 Battery Replacement: step 12, image 3 of 3
    • Use the plastic label to remove the battery from its recess.

kyle2

Member since: 06/03/10

101 Reputation

1 Guide authored

Team

Team Freedom Member of Team Freedom

Community

1 Member

3 Guides authored

View Statistics:

Past 24 Hours: 0

Past 7 Days: 4

Past 30 Days: 33

All Time: 2,757