Siemens-Maquet Servo-i Filter for the Inspiration Pressure Transducer Replacement

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.

What you need

  1. Siemens-Maquet Servo-i Filter for the Inspiration Pressure Transducer Replacement, Filter for Gas Modules: step 1, image 1 of 1
    • When performing this maintenance, a 'Maintenance kit, 5,000 hours' should be used

    • Remove the inspiratory section cover

    • Loosen the screw (1)

    • Lift off the gas module bracket (2)

    • Service Manual 4 - 14

  2. Siemens-Maquet Servo-i Filter for the Inspiration Pressure Transducer Replacement: step 2, image 1 of 1
    • Pull out and lift off both gas modules.

    • Service Manual 4 - 14

  3. Siemens-Maquet Servo-i Filter for the Inspiration Pressure Transducer Replacement: step 3, image 1 of 1
    • Note 1: When assembling, make sure that the connector muff properly seals around the inspiratory pipe (1). The connector muff must not be pushed too far onto the nozzle units (2)

    • Note 2: The gas modules used in Servo-i are factory adjusted for this purpose. When replacing gas modules, make sure to use only Servo-i gas modules.Similar gas modules intended for the SV 300/300A or KION / KION-i Systems must not be used.

    • Service Manual 4 - 14

  4. Siemens-Maquet Servo-i Filter for the Inspiration Pressure Transducer Replacement: step 4, image 1 of 1
    • Replace filters in the gas modules. When replacing filter, move the rubber seal from the old to the new filter.

    • Service Manual 7 - 3

  5. Siemens-Maquet Servo-i Filter for the Inspiration Pressure Transducer Replacement, Filter for the Inspiration Pressure Transducer: step 5, image 1 of 1
    • Locate the inspiratory pressure transducer filter shown by C in the image.

  6. Siemens-Maquet Servo-i Filter for the Inspiration Pressure Transducer Replacement: step 6, image 1 of 1
    • Replace the filter for the inspiratory pressure transducer.

    • Service Manual 7 - 4

  7. Siemens-Maquet Servo-i Filter for the Inspiration Pressure Transducer Replacement: step 7, image 1 of 1
    • Make sure that the filter is correctly seated into the rubber ring.

    • Service Manual 7 - 4

    • Note: This filter may already be replaced as a part of the 'Extended cleaning of Inspiratory channel' performed in combination with the Preventive maintenance.

    • Service Manual 7 - 4