Skip to main content

How to Disconnect a Philips Respironics V60

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.
  1. How to Disconnect a Philips Respironics V60, shut down the ventilator: step 1, image 1 of 2 How to Disconnect a Philips Respironics V60, shut down the ventilator: step 1, image 2 of 2
    • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.

    • Select Ventilator Shutdown, which will shut down the ventilator.

    • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

    • Disconnect the ventilator from the AC power.

    • Unplug the ventilator

  2. How to Disconnect a Philips Respironics V60: step 4, image 1 of 1
    • Disconnect the internal battery from the power harness connector.

Conclusion

To reassemble your device, follow these instructions in reverse order.