Tektronix TBS1062 Oscilloscope Disassembly

What you need

  1. Tektronix TBS1062 Oscilloscope Disassembly, Removing the front panel: step 1, image 1 of 3 Tektronix TBS1062 Oscilloscope Disassembly, Removing the front panel: step 1, image 2 of 3 Tektronix TBS1062 Oscilloscope Disassembly, Removing the front panel: step 1, image 3 of 3
    • Remove all eight plastic buttons.

    • Remove the five Torx T15 screws.

    • Remove the front panel by pulling it off

  2. Tektronix TBS1062 Oscilloscope Disassembly, Removing the back cover: step 2, image 1 of 2 Tektronix TBS1062 Oscilloscope Disassembly, Removing the back cover: step 2, image 2 of 2
    • Remove the power plug from the cover

    • Firmly grasp the ON/OFF switch and pull it straight out of the box.

    • Press the button into the ON/OFF button recess until it clicks into place.

    • Remove the two black plastic plugs from the bottom.

    • Remove the five Torx T15 screws.

    • Remove the back cover.

  3. Tektronix TBS1062 Oscilloscope Disassembly, Power supply removal - ATTENTION: step 3, image 1 of 3 Tektronix TBS1062 Oscilloscope Disassembly, Power supply removal - ATTENTION: step 3, image 2 of 3 Tektronix TBS1062 Oscilloscope Disassembly, Power supply removal - ATTENTION: step 3, image 3 of 3
    • Even when disconnected, the power supply can retain a dangerously high voltage. Avoid contact with the rear of the board or other components. Use insulated electrician's gloves. Ensure that the capacitors are discharged before touching them.

    • Disconnect the J101 power supply connector from the motherboard.

    • Disconnect the earth connection.

    • Remove the 5x Torx T15 screws.

    • Remove the power supply board.

  4. Tektronix TBS1062 Oscilloscope Disassembly, Control panel removal: step 4, image 1 of 2 Tektronix TBS1062 Oscilloscope Disassembly, Control panel removal: step 4, image 2 of 2
    • Disconnect the black cable from the motherboard

    • Remove the 5x Torx T15 screws

    • Pull the control board plate

  5. Tektronix TBS1062 Oscilloscope Disassembly, Removing the mainboard: step 5, image 1 of 3 Tektronix TBS1062 Oscilloscope Disassembly, Removing the mainboard: step 5, image 2 of 3 Tektronix TBS1062 Oscilloscope Disassembly, Removing the mainboard: step 5, image 3 of 3
    • Loosen the three nuts holding the signal connectors.

    • Loosen the three T15 Torx screws.

    • Remove the cables.

    • Remove the main board.

  6. Tektronix TBS1062 Oscilloscope Disassembly, LCD display removal: step 6, image 1 of 3 Tektronix TBS1062 Oscilloscope Disassembly, LCD display removal: step 6, image 2 of 3 Tektronix TBS1062 Oscilloscope Disassembly, LCD display removal: step 6, image 3 of 3
    • Disconnect the ribbon cable and the LCD backlight cable from the motherboard.

    • Loosen the 4x 8mm nuts from the display supports.

    • Remove the LCD display.

Conclusion

To reassemble your device, follow these instructions in reverse order.

Special thanks to these translators:

pt en

100%

These translators are helping us fix the world! Want to contribute?
Start translating ›

Jose Gustavo Abreu Murta

Member since: 07/06/10

4,598 Reputation

9 Guides authored

2 Comments

This guide is written in English, not Portugese. How can that be changed?

Christopher Sean Harris - Reply

Hi Christofer.

Do you intend to translate into Portuguese? I messed up the language when editing it.

Jose Gustavo Abreu Murta - Reply

Add Comment

View Statistics:

Past 24 Hours: 0

Past 7 Days: 1

Past 30 Days: 14

All Time: 386