Skip to main content

Free Shipping on Domestic Orders $75+

My ophthalmoscope is not charging

When I attach my opgthalmoscope to the charging dock it keeps on flashing lights fromcthe globe head but does not charge or turns on when I use it

Answer this question I have this problem too

Is this a good question?

Score 1
Add a comment

1 Answer

1️⃣ Most Common Cause – Dead / Failed Battery

If the globe head flashes when docked but won’t power on:

  • The rechargeable handle battery may be completely dead.
  • It may no longer hold charge.
  • It could be over-discharged.

Fix:
Replace the rechargeable battery inside the handle.
(These typically use Welch Allyn rechargeable NiMH or Li-ion handle batteries depending on model.)

2️⃣ Charging Contacts Dirty or Oxidized

If the light flashes but no charging happens:

  • Check metal charging contacts on:
    • The bottom of the handle
    • Inside the charging dock

Fix:
Clean with:

  • Isopropyl alcohol (70%+)
  • Cotton swab
    Let dry fully before reconnecting.

3️⃣ Faulty Charging Dock

Even if it lights up, the dock may not be delivering proper voltage.

Check:

  • Try another compatible Welch Allyn charger if available.
  • Use a multimeter to test dock output voltage.

4️⃣ Handle Internal Connection Loose

If flashing happens only when slightly moved:

  • Internal wiring in the handle may be loose.
  • The battery terminal spring may not be making contact.

This requires opening the handle carefully.

5️⃣ Protection Lock / Reset

Some Welch Allyn rechargeable handles need:

  • 8–12 hours full dock charging after full drain
  • Proper seating (twist firmly into dock)

Try:

  1. Remove battery.
  2. Reinsert firmly.
  3. Place in dock overnight (minimum 8 hours).

Was this answer helpful?